Catalase (CAT) Activity Assay Kit

To better serve you, we would like to discuss your specific requirement, Please Contact Us for a quote.

Catalase (CAT) Activity Assay Kit

SKU # E-BC-K031-S
Detection Instrument Spectrophotometer(405 nm)
Detection Method Colorimetric method


Product Details

Properties

Synonyms CAT
Sample Type Serum, plasma, cells, tissue
Sensitivity 0.27 U/mL
Detection Range 0.27-155.4 U/mL
Detection Method Colorimetric method
Assay type Enzyme Activity
Assay time 70 min
Precision Average inter-assay CV: 5.100% | Average intra-assay CV: 3.100%
Other instruments required Micropipettor, Vortex mixer, Incubator or Water-bath
Other reagents required Normal saline (0.9% NaCl) or PBS (0.01 M, pH 7.4)
Storage 2-8℃
Valid period 12 months


Dilution of Sample

It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (0.27-155.4 U/mL).

The recommended dilution factor for different samples is as follows (for reference only):

Sample type Dilution factor The volume of sample
10% Rat liver tissue homogenization 25-50 50 μL
10% Rat kidney tissue homogenization 10-25 50 μL
10% Rat brain tissue homogenization 05-10 50 μL
Human serum 1 100 μL
293T supernatant 1 100 μL

 

Note: The diluent is normal saline (0.9% NaCl) or PBS (0.01 M, pH 7.4).

 

Detection Principle

The reaction that catalase (CAT) decomposes H2O2 can be quickly stopped by ammonium molybdate. The residual H2O2 reacts with ammonium molybdate to generate a yellowish complex. CAT activity can be calculated by production of the yellowish complex at 405 nm.


Kit Components & Storage

Item Component Size 1 (50 assays) Size 2 (100 assays) Storage
Reagent 1 Buffer Solution 60 mL × 1 vial 60 mL × 2 vials 2-8℃, 12 months
Reagent 2 Substrate 12 mL × 1 vial 12 mL × 1 vial 2-8℃, 12 months
Reagent 3 Chromogenic Agent Powder ×1 vial Powder × 2 vials 2-8℃, 12 months
Reagent 4 Clarificant 6 mL × 1 vial 12 mL × 1 vial 2-8℃, 12 months

 

Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

Technical Data:

Parameter:

Intra-assay Precision

Three human serum samples were assayed in replicates of 20 to determine precision within an assay (CV = Coefficient of Variation).

Parameters Sample 1 Sample 2 Sample 3
Mean (U/mL) 3.50 34.50 88.50
%CV 3.6 3.2 2.5


Inter-assay Precision

Three human serum samples were assayed 20 times in duplicate by three operators to determine precision between assays.

Parameters Sample 1 Sample 2 Sample 3
Mean (U/mL) 3.50 34.50 88.50
%CV 5.7 4.9 4.7


Recovery

Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 96%


Sample 1 Sample 2 Sample 3
Expected Conc. (U/mL) 19 77 106
Observed Conc. (U/mL) 18.4 73.2 101.8
Recovery rate (%) 97 95 96


Sensitivity

The analytical sensitivity of the assay is 0.27 U/mL CAT. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.