Thank you!

Your quote has been successfully submitted!

For products requiring additional information, our team will contact you within 1 business day

Failed

There was an error submitting your quote. Please try again.

Ethanol Colorimetric Assay Kit– MSE Supplies LLC

Free Shipping on MSE PRO Online Orders of $500 or More! U.S. Orders Only * Offer Excludes Hazmat Shipments *

Menu

This product has been added to the cart.

Ethanol Colorimetric Assay Kit

SKU: E-BC-K891-M-500

  • $ 93595



Ethanol Colorimetric Assay Kit

SKU # E-BC-K891-M
Detection Instrument Microplate reader (440-460 nm, optimum wavelength: 450 nm)
Detection method
Colorimetric method


Product Details

Properties

Sample type Serum, plasma, wine
Sensitivity 0.27 μmol/mL
Detection range 0.27–10.0 μmol/mL
Detection method Colorimetric method
Assay type Quantitative
Assay time 30 min
Precision Average inter-assay CV: 5% | Average intra-assay CV: 3.500%
Other instruments required Micropipettor, Vortex mixer, Centrifuge, Incubator (37℃)
Storage -20℃
Valid period 12 months

 

Dilution of Sample

 

It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (0.27–17.0 μmol/mL).

The recommended dilution factor for different samples is as follows (for reference only):

Sample type Dilution factor
Beer (3.6% alcohol) 40-80
White wine (12% alcohol) 250-300


Note: The diluent is double distilled water.

 

Detection Principle

Ethanol dehydrogenase can catalyze oxidative dehydrogenation of ethanol to acetaldehyde, and NAD+ is reduced to produce NADH. NADH, under the action of 1-mPMS, transfer electrons to WST-8 to produce the yellow product, which has a characteristic absorption peak at 450 nm. Therefore, ethanol content can be quantified by measure the OD value at 450 nm.

 

Kit Components & Storage

Item Component Size 1(48 T) Size 2(96 T) Storage
Reagent 1 Buffer Solution A 10 mL × 1 vial 20 mL × 1 vial -20℃, 12 months
Reagent 2 Enzyme Reagent Powder × 1 vial Powder × 2 vials -20℃, 12 months, shading light
Reagent 3 Buffer Solution B 7 mL × 1 vial 14 mL × 1 vial -20℃, 12 months
Reagent 4 Substrate Powder × 1 vial Powder × 2 vials -20℃, 12 months, shading light
Reagent 5 Chromogenic Agent 1.5 mL × 1 vial 1.5 mL × 2 vials -20℃, 12 months, shading light
Reagent 6 10 μmol/mL Standard Solution 1.8 mL × 1 vial 1.8 mL × 2 vials -20℃, 12 months
Microplate 96 wells No requirement
Plate Sealer 2 pieces

 

Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

Technical Data:

Parameter:

Intra-assay Precision

Three beer samples were assayed in replicates of 20 to determine precision within an assay (CV = Coefficient of Variation).

Parameters Sample 1 Sample 2 Sample 3
Mean (μmol/mL) 0.80 2.60 6.80
%CV 3.8 3.5 3.2


Inter-assay Precision

Three beer samples were assayed 20 times in duplicate by three operators to determine precision between assays

Parameters Sample 1 Sample 2 Sample 3
Mean (μmol/mL) 0.80 2.60 6.80
%CV 4.8 5.1 5.1


Recovery

Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 96%.


Standard 1 Standard 2 Standard 3
Expected Conc. (μmol/mL) 2.7 5 7.5
Observed Conc. (μmol/mL) 2.6 4.7 7.2
Recovery rate(%) 98 94 96

 

Sensitivity

The analytical sensitivity of the assay is 0.27 μmol/mL. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.

 

Standard Curve

As the OD value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only:

Concentration (μmol/mL) 0 2 3 4 6 8 9 10
Average OD 0.016 0.147 0.212 0.280 0.416 0.550 0.601 0.669