Heidolph Tuttnauer Lab Line Vertical Autoclave 5075ELVC
SKU: 023210990
Heidolph Tuttnauer Lab Line Vertical Autoclave 5075ELVC
The 5075ELVC model sterilizer comes with a 50cm chamber diameter size with a depth of 75cm.
The unit comes standard with an internal thermometer, water sensing electrode, external pressure gauge and a printer.
This unit runs at 3-phase 230V and comes with a set of 2 large wire baskets.
Compressed air can be connected to this unit to use its cooling function.
Product Parameters
Chamber Volume | 106 L |
Mobile is not an option and must be connected to building utilities for distilled water, tap water and drain | |
Chamber Dimensions (Dia. x Depth) | 50 x 75 cm |
Overall Dimensions (D x W x H) | 110 x 86.6 x 71.5 cm |
43 x 34 x 28 in | |
Max Temperature | 135C |
Chamber Material | Stainless Steel 316 Ti (1.4571) |
Housing/Door Material | Stainless Steel 304 L |
Internal Thermometer | included |
Sample Temperature Probes | set of 2, included |
Water-sensing Electrode | included |
External Pressure Gauge | included |
Printer | included |
USB | included |
Wire Baskets | set of 2 (large) |
Wheels w/lock | n/a |
Customer Must Supply | These units do not come with a plug. Nema (L15-30P & R of L14-30R & P) will need to be installed |
Voltage Requirements | 230 V / 50-60 Hz, Single Phase |
Amps | 30A |
Power Requirements | 9.0 KW |
Compressed Air | 1/2" 3-4 Bar (44-58 psi) |
Tap Water | 1/2" 3-4 Bar (29-44 psi) |
Mineral Free Water | 1/2" 3-4 Bar (29-44 psi) RO or Distilled |
Drain | 2" Min Withstanding temp of 80C (176F) |