Human NT-4 (Neurotrophin 4) CLIA Kit
SKU: E-CL-H1172
Human NT-4 (Neurotrophin 4) CLIA Kit
Detection Range | 31.25-2000 pg/mL |
Sensitivity | 18.75 pg/mL |
Product Details
Properties
Assay type | Sandwich-CLIA |
Format | 96T/48T |
Assay time | 3.5h |
Detection range | 31.25-2000 pg/mL |
Sensitivity | 18.75 pg/mL |
Sample type &Sample volume | serum, plasma and other biological fluids; 100μL |
Specificity | This kit recognizes Human NT-4 in samples. No significant cross-reactivity or interference between Human NT-4 and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are < 15%. |
Application | This CLIA kit applies to the in vitro quantitative determination of Human NT-4 concentrations in serum, plasma and other biological fluids. |
Test Principle
Kit Components & Storage
An unopened kit can be stored at 2-8℃ for 1 month. If the kit is not supposed to be used within 1 month, store the components separately according to the following conditions once the kit is received.
Item | Specifications | Storage |
---|---|---|
Micro CLIA Plate(Dismountable) | 96T: 8 wells ×12 strips 48T: 8 wells ×6 strips |
-20℃, 12 months |
Reference Standard | 96T: 2 vials 48T: 1 vial |
|
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 48T: 1 vial, 60 μL |
|
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 48T: 1 vial, 60 μL |
-20℃(Protect from light), 12 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 12 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | |
HRP Conjugate Diluent | 1 vial, 14 mL | |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃(Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy |
Technical Data:
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Human NT-4 were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Human NT-4 were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 105.60 | 267.46 | 841.29 | 115.36 | 251.07 | 925.33 |
Standard deviation | 9.22 | 24.50 | 67.64 | 11.95 | 29.00 | 70.60 |
CV (%) | 8.73 | 9.16 | 8.04 | 10.36 | 11.55 | 7.63 |
Recovery
The recovery of Human NT-4 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
---|---|---|
Serum(n=8) | 97-109 | 104 |
EDTA plasma(n=8) | 99-117 | 106 |
Cell culture media(n=8) | 85-99 | 92 |
Assay Procedures
1.Add 100 μL of standard or sample to each well. Incubate for 90 min at 37℃. |
|
2.Remove the liquid. |
|
3.Add 100 μL Biotinylated Detection Ab. Incubate 1 hour at 37℃. Aspirate and wash 3 times. |
|
4.Add 100 μL HRP Conjugate. Incubate 30 min at 37℃. Aspirate and wash 5 times. |
|
5.Add 100 μL Substrate Mixture Solution. Incubate for 5 min at 37℃. |
|
6.Fluorescence appeared. Measure the RLU value with the Chemiluminescence immunoassay analyzer. Calculation of the results. |