Magnesium (Mg) Colorimetric Assay Kit– MSE Supplies LLC

Free Shipping on MSE PRO Online Orders of $500 or More! U.S. Orders Only * Offer Excludes Hazmat Shipments *

Menu

Magnesium (Mg) Colorimetric Assay Kit

SKU: E-BC-K162-S-100

To better serve you, we would like to discuss your specific requirement.
Please Contact Us for a quote.

Magnesium (Mg) Colorimetric Assay Kit

SKU # E-BC-K162-S
Detection Instrument Spectrophotometer (540 nm)
Detection method Colorimetric method


Product Details

Properties

Synonyms Mg
Sample Type Serum, plasma
Sensitivity 0.12 mmol/L
Detection Range 0.12-2.50 mmol/L
Detection Method Colorimetric method
Assay type Quantitative
Assay time 50 min
Precision Average inter-assay CV: 9.900% | Average intra-assay CV: 4.800%
Other instruments required Micropipettor, Centrifuge, Incubator, Vortex mixer
Other reagents required Normal saline (0.9% NaCl), PBS (0.01 M, pH 7.4)
Storage 2-8℃
Valid period 12 months


Images

A Raja et al investigate the anti-osteoporotic effect of Matricaria chamomilla L. Flower. Magnesium (Mg) level in rat serum was determined using Mg colorimetric assay kit (E-BC-K162-S).

The levels of Mg concentration was significantly changed comparing to the diseased control. (**P<0.01, ***P<0.001)

 

Dilution of Sample

It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (0.12-2.50 mmol/L).

The recommended dilution factor for different samples is as follows (for reference only):

Sample type Dilution factor
Human serum 1
Rat serum 1
Mouse serum 1
Porcine serum 1
Chicken serum 1
Dog serum
1

 

Note: The diluent is normal saline (0.9% NaCl) or PBS (0.01 M, pH 7.4).

 

Detection Principle

The magnesium in the serum reacts with the complexometric indicator (Calmagite) to form the Calmagite-Mg compound. The absorbance of this compound at 540 nm is proportional to the concentration of magnesium in the sample. The concentration of magnesium can be calculated by measuring the OD value at 540 nm.

 

Kit Components & Storage

Item Component Size 1
(50 assays)
Size 2
(100 assays)
Storage
Reagent 1 Alkali Reagent 30 mL × 1 vial 60 mL × 1 vial 2-8°C, 12 months
Reagent 2 Chromogenic Agent 30 mL × 1 vial 60 mL × 1 vial 2-8°C, 12 months,
shading light
Reagent 3 5 mmol/L Magnesium Standard 1 mL × 1 vial 1 mL × 2 vials 2-8°C, 12 months

 

Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

Technical Data:

Parameter:

Intra-assay Precision

Three human serum samples were assayed in replicates of 20 to determine precision within an assay. (CV = Coefficient of Variation)

Parameters Sample 1 Sample 2 Sample 3
Mean (mmol/L) 0.55 1.35 2.30
%CV 5.3 4.6 4.5


Inter-assay Precision

Three human serum samples were assayed 20 times in duplicate by three operators to determine precision between assays.

Parameters Sample 1 Sample 2 Sample 3
Mean (mmol/L) 0.55 1.35 2.30
%CV 10.2 9.4 10.1

 

Recovery

Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 95%.

Standard 1 Standard 2 Standard 3
Expected Conc. (mmol/L) 0.35 1.06 1.95
Observed Conc. (mmol/L) 0.3 1.0 1.8
Recovery rate (%) 94 98 93

 

Sensitivity

The analytical sensitivity of the assay is 0.12 mmol/L. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.

 

Standard Curve

As the OD value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only.

Concentration (mmol/L) 0 0.5 0.75 1 1.5 2 2.5
Average OD 0.331 0.374 0.390 0.410 0.445 0.484 0.514
Absoluted OD 0 0.043 0.059 0.079 0.114 0.153 0.183