Thank you!

Your quote has been successfully submitted!

For products requiring additional information, our team will contact you within 1 business day

Failed

There was an error submitting your quote. Please try again.

Malondialdehyde (MDA) Colorimetric Assay Kit (Plant Samples)– MSE Supplies LLC

Free Shipping on MSE PRO Online Orders of $500 or More! U.S. Orders Only * Offer Excludes Hazmat Shipments *

Menu

Malondialdehyde (MDA) Colorimetric Assay Kit (Plant Samples)

SKU: E-BC-K027-M-96

To better serve you, we would like to discuss your specific requirement.
Please Contact Us for a quote.

Malondialdehyde (MDA) Colorimetric Assay Kit (Plant Samples)

SKU # E-BC-K027-M
Detection Instrument Microplate reader(525-535 nm)
Detection Method Colorimetric method


Product Details

Properties

Synonyms MDA
Sample Type Plant tissue
Sensitivity 0.17 nmol/mL
Detection Range 0.17-50 nmol/mL
Detection Method Colorimetric method
Assay type Quantitative
Assay time 65 min
Precision Average inter-assay CV: 5.100% | Average intra-assay CV: 4.500%
Other instruments required Micropipettor, Vortex mixer, Incubator, Centrifuge
Other reagents required Absolute ethanol
Storage 2-8℃
Valid period 12 months


Detection Principle

MDA in the catabolite of lipid peroxide can react with thiobarbituric acid (TBA) and produce red compound, which has a maximum absorption peak at 532 nm.

 

Kit Components & Storage

Item Component Size 1(48 T) Size 2(96 T) Storage
Reagent 1 Clarificant 1.5 mL × 1 vial 1.5 mL × 2 vials 2-8℃, 12 months
Reagent 2 Acid Reagent 45 mL × 1 vial 45 mL × 2 vials 2-8℃, 12 months
Reagent 3 Chromogenic Agent 15 mL × 1 vial 30 mL × 1 vial 2-8℃, 12 months shading light
Reagent 4 200 nmol/mL Standard 5 mL × 1 vial 5 mL × 1 vial 2-8℃, 12 months
Reagent 5 10 × Concentrated Extracting Solution 40 mL × 1 vial 40 mL × 1 vial 2-8℃, 12 months
Microplate 96 wells No requirement
Plate Sealer 2 pieces

 

Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

Technical Data:

Parameter:

Intra-assay Precision

Three carrot samples were assayed in replicates of 20 to determine precision within an assay (CV = Coefficient of Variation).

Parameters Sample 1 Sample 2 Sample 3
Mean (nmol/mL) 0.50 15.50 35.60
%CV 4.9 4.5 4.4


Inter-assay Precision

Three carrot samples were assayed 17 times in duplicate by three operators to determine precision between assays

Parameters Sample 1 Sample 2 Sample 3
Mean (nmol/mL) 0.50 15.50 35.60
%CV 6.2 6.7 6.3


Recovery

Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 100%.


Sample 1 Sample 2 Sample 3
Expected Conc. (nmol/mL) 8.5 18.3 33.5
Observed Conc. (nmol/mL) 8.6 18.1 33.5
Recovery rate (%) 101 99 100


Sensitivity

The analytical sensitivity of the assay is 0.38 nmol/mL MDA. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.

 

Standard Curve:

As the OD value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only:

Concentration (nmol/mL) 0 5 10 10 20 30 40 50
Average OD 0.040 0.138 0.242 0.333 0.442 0.665 0.878 1.014
Absoluted OD 0 0.098 0.202 0.293 0.402 0.625 0.838 0.974