Thank you!

Your quote has been successfully submitted!

For products requiring additional information, our team will contact you within 1 business day

Failed

There was an error submitting your quote. Please try again.

Malondialdehyde (MDA) Fluorometric Assay Kit– MSE Supplies LLC

Free Shipping on MSE PRO Online Orders of $500 or More! U.S. Orders Only * Offer Excludes Hazmat Shipments *

Menu

This product has been added to the cart.

Malondialdehyde (MDA) Fluorometric Assay Kit

SKU: E-BC-F007-96

  • $ 27295


Malondialdehyde (MDA) Fluorometric Assay Kit

SKU # E-BC-F007
Detection Instrument Fluorescence microplate reader (Ex/Em=520 nm/550 nm)
Detection Method Fluorescence method


Product Details

Properties

Synonyms MDA
Sample type Serum, plasma, animal tissue, cell
Sensitivity 0.04 μmol/L
Detection Range 0.04-10 μmol/L
Detection Method Fluorescence method
Precision Inter-assay CV: 4.5-7.0% | intra-assay CV: 2.3-4.3%
Storage This product can be stored at 2~8°C for 12 months with shading light.
Valid period 12 months

 

Dilution Of Sample

The recommended dilution factor for different samples is as follows (for reference only):

Sample type Dilution factor
Human serum 2-4
Human plasma 1
Rat serum 1
Rat plasma 1
Mouse serum 2-4
Mouse plasma 1
10% Mouse brain tissue homogenate 1
10% Rat brain tissue homogenate 1
10% Mouse liver tissue homogenate 1
10% Rat kidney tissue homogenate 2-4
9.2×10^6 CHO cells 1

 

The diluent of serum (plasma) or animal tissue is normal saline (0.9% NaCl) or PBS (0.01 M, pH 7.4); The diluent of cells is CAMT lysis buffer. For the dilution of other sample types, please do pretest to confirm the dilution factor.

 

Detection Principle

MDA in the catabolite of lipid peroxide can react with thiobarbituric acid (TBA) and produce red compound, which is proportional to the fluorescence intensity at the excitation wavelength of 520 nm and emission wavelength of 550 nm.


Kit Components & Storage

Item Component Size 1(48 T) Size 2(96 T) Storage
Reagent 1 Clarificant 6 mL ×1 vial 12 mL × 1 vial 2-8°C, 12 months
Reagent 2 Acid Reagent 2 mL ×1 vial 4 mL × 1 vial 2-8°C, 12 months
Reagent 3 TBA Reagent Powder ×1 vial Powder × 1 vial 2-8°C, 12 months shading light
Reagent 4 20 μmol/L Standard 5 mL ×1 vial 5 mL × 1 vial 2-8°C, 12 months
Reagent 5 CAMT Lysis Buffer 20 mL ×1 vial 40 mL × 1 vial 2-8°C, 12 months
Black Microplate 96 wells No requirement
Plate Sealer 2 pieces


Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.


    Technical Data:

    Parameter:

    Intra-assay Precision

    Three human serum samples were assayed in replicates of 20 to determine precision within an assay. (CV = Coefficient of Variation)

    Parameters Sample 1 Sample 2 Sample 3 
    Mean (μmol/L) 0.80 4.50 7.30
    %CV 4.3 3.3 2.3

     

    Inter-assay Precision

    Three human serum samples were assayed 20 times in duplicate by three operators to determine precision between assays.

    Parameters Sample 1 Sample 2 Sample 3 
    Mean (μmol/L) 0.80 4.50 7.30
    %CV 4.5 7.0 5.6

     

    Recovery

    Take three samples of high concentration, middle concentration and low
    concentration to test the samples of each concentration for 6 times parallelly to get
    the average recovery rate of 97%.


    Standard 1 Standard 2 Standard 3
    Expected Conc.(μmol/L) 0.7 3.5 7.5
    Observed Conc.(μmol/L) 0.7 3.5 7.3
    Recovery Rate (%) 95 99 97

     

    Sensitivity

    The analytical sensitivity of the assay is 0.04 μmol/L. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.

     

    Standard Curve:

    As the fluorescence value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only:

    Concentration (μmol/L) 0 0.5 1 2 4 6 8 10
    Fluorescence value 57 164 262 486 893 1339 1780 2105
    62 168 258 495 924 1373 1757 2083
    Average fluorescence value 59 166 260 490 909 1356 1768 2094
    Absoluted fluorescence value 0 107 1 431 850 1297 1709 2035