QuicKey Pro Mouse IgG3(Immunoglobulin G3) ELISA Kit– MSE Supplies LLC

Free Shipping on MSE PRO Online Orders of $500 or More! U.S. Orders Only * Offer Excludes Hazmat Shipments *

Menu

This product has been added to the cart.

QuicKey Pro Mouse IgG3(Immunoglobulin G3) ELISA Kit

SKU: E-OSEL-M0009

  • $ 44295


QuicKey Pro Mouse IgG3(Immunoglobulin G3) ELISA Kit

Detection Range 0.31-20 ng/mL
Sensitivity 0.14 ng/mL



Product Details

Properties

Assay type Sandwich-ELISA
Format 96T
Assay time 1.5h
Detection range 0.31-20 ng/mL
Sensitivity 0.14 ng/mL
Sample type &Sample volume serum, plasma and other biological fluids; 50μL
Specificity This kit recognizes IgG3 in samples. No significant cross-reactivity or interference between IgG3 and analogues was observed.
Reproducibility Both intra-CV and inter-CV are < 10%.
Application This ELISA kit applies to the in vitro quantitative determination of IgG3 concentrations in serum, plasma and other biological fluids.

 

Test Principle

This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Mouse IgG3. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Mouse IgG3 are added to the micro ELISA plate wells. Mouse IgG3 in samples (or standards) combines with the coated antibody and HRP linked detection antibody specific to IgG3. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The concentration of Mouse IgG3 in the samples is then determined by comparing the OD of the samples to the standard curve.

 

Kit Components and Storage

An unopened kit can be stored at 2-8℃ for 6 months. After test, the unused wells and reagents should be stored according to the table.

Item Specifications Storage
Micro ELISA Plate(Dismountable) 96T: 8 wells ×12 strips
48T: 8 wells ×6 strips
24T: 8 wells ×3 strips
96T*5: 5 plates, 96T
2-8℃, 1 months
Reference Standard 96T: 2 vials
48T/24T: 1 vial
96T*5: 10 vials
2-8℃, use the reconstituted standard within 24h
Concentrated HRP Linked Detection Ab(100×) 96T: 1 vial, 60 μL
48T/24T: 1 vial, 30 μL
96T*5: 5 vials, 60 μL
2-8℃(Protect from light)
Reference Standard & Sample Diluent 96T/48T/24T: 1 vial, 20 mL
96T*5: 5 vials, 20 mL
2-8℃
HRP Linked Ab Diluent 96T/48T/24T: 1 vial, 14 mL
96T*5: 5 vials, 14 mL
Concentrated Wash Buffer(25×) 96T/48T/24T: 1 vial, 30 mL
96T*5: 5 vials, 30 mL
Substrate Reagent 96T/48T/24T: 1 vial, 10 mL
96T*5: 5 vials, 10 mL
2-8℃(Protect from light)
Stop Solution 96T/48T/24T: 1 vial, 10 mL
96T*5: 5 vials, 10 mL
2-8℃
Plate Sealer 96T/48T/24T: 5 pieces
96T*5: 25 pieces
Product Description 1 copy
Certificate of Analysis 1 copy

 

Technical Data:

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level IgG3 were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level IgG3 were tested on 3 different plates, 20 replicates in each plate.

  Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (ng/mL) 1.02 2.84 7.14 1.09 3.02 6.65
Standard deviation 0.07 0.16 0.29 0.06 0.18 0.27
CV (%) 6.77 5.46 4.08 5.83 5.90 4.02

 

 Recovery

The recovery of IgG3 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range (%) Average Recovery (%)
Serum(n=8) 86-101 92
EDTA plasma(n=8) 91-101 96
Cell culture media(n=8) 93-108 101


Target Information

Database Links     SwissProt: IgG3
Synonyms     50μL
Research Area     Immunology


Assay Procedures

1. Add 50µL standard or sample to the wells, Immediately add 50µL of HRP linked Detection Ab working solution to each well. Incubate for 60 min at 37°C.
2. Aspirate and wash the plate for 5 times.
3. Add 90µL of Substrate Reagent. Incubate for about 15 min at 37°C.
4. Add 50µL Stop Solution.
5. Read the plate at 450nm immediately. Calculation of the results