Thank you!

Your quote has been successfully submitted!

For products requiring additional information, our team will contact you within 1 business day

Failed

There was an error submitting your quote. Please try again.

Rat Pg(Progesterone) ELISA Kit– MSE Supplies LLC

Free Shipping on MSE PRO Online Orders of $500 or More! U.S. Orders Only * Offer Excludes Hazmat Shipments *

Menu

This product has been added to the cart.

Rat Pg(Progesterone) ELISA Kit

SKU: E-OSEL-R0007

  • $ 44295



Rat Pg(Progesterone) ELISA Kit

Detection Range 15.63-1000 pg/mL
Sensitivity 7.01 pg/mL



Product Details

Properties

Assay type Competitive-ELISA
Format 96T
Assay time 1.5h
Detection range 15.63-1000 pg/mL
Sensitivity 7.01 pg/mL
Sample type &Sample volume serum and plasma; 50μL
Specificity This kit recognizes Pg in samples. No significant cross-reactivity or interference between Pg and analogues was observed.
Reproducibility Both intra-CV and inter-CV are < 10%.
Application This ELISA kit applies to the in vitro quantitative determination of Pg concentrations in serum and plasma.

 

Test Principle

This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Rat Pg. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Rat Pg are added to the micro ELISA plate wells. Rat Pg in samples (or standards) competes with a fixed amount of Pg on the solid phase supporter for sites on the HRP linked detection antibody specific to Pg. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The concentration of Rat Pg in the samples is then determined by comparing the OD of the samples to the standard curve.

 

Kit Components and Storage

An unopened kit can be stored at 2-8℃ for 6 months. After test, the unused wells and reagents should be stored according to the table.

Item Specifications Storage
Micro ELISA Plate(Dismountable) 96T: 8 wells ×12 strips
48T: 8 wells ×6 strips
24T: 8 wells ×3 strips
96T*5: 5 plates, 96T
2-8℃, 1 months
Reference Standard 96T: 2 vials
48T/24T: 1 vial
96T*5: 10 vials
2-8℃, use the reconstituted standard within 24h
Concentrated HRP Linked Detection Ab(100×) 96T: 1 vial, 60 μL
48T/24T: 1 vial, 30 μL
96T*5: 5 vials, 60 μL
2-8℃(Protect from light)
Reference Standard & Sample Diluent 96T/48T/24T: 1 vial, 20 mL
96T*5: 5 vials, 20 mL
2-8℃
HRP Linked Ab Diluent 96T/48T/24T: 1 vial, 14 mL
96T*5: 5 vials, 14 mL
Concentrated Wash Buffer(25×) 96T/48T/24T: 1 vial, 30 mL
96T*5: 5 vials, 30 mL
Substrate Reagent 96T/48T/24T: 1 vial, 10 mL
96T*5: 5 vials, 10 mL
2-8℃(Protect from light)
Stop Solution 96T/48T/24T: 1 vial, 10 mL
96T*5: 5 vials, 10 mL
2-8℃
Plate Sealer 96T/48T/24T: 5 pieces
96T*5: 25 pieces
Product Description 1 copy
Certificate of Analysis 1 copy

 

Technical Data:

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Pg were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Pg were tested on 3 different plates, 20 replicates in each plate.

  Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (pg/mL) 51.82 153.05 376.37 50.57 167.62 390.77
Standard deviation 3.26 7.61 17.39 2.91 8.55 19.58
CV (%) 6.30 4.97 4.62 5.76 5.10 5.01

 

 Recovery

The recovery of Pg spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range (%) Average Recovery (%)
Serum(n=8) 91-101 96
EDTA plasma(n=8) 94-107 101


Target Information

Synonyms     Progesterone,Pg
Research Area     Signal Transduction


Assay Procedures

1. Add 50µL each standard and sample. Immediately add 50µL of HRP linked Ab working solution. Incubate for 60 min at 37°C.
2. Aspirate and wash the plate for 5 times.
3. Add 90µL of Substrate Reagent. Incubate for about 15 min at 37°C.
4. Add 50µL of Stop Solution.
5. Read the plate at 450nm immediately. Calculation of the results