Vitamin E (VE) Colorimetric Assay Kit
SKU: E-BC-K033-S-100
To better serve you, we would like to discuss your specific requirement.
Please Contact Us for a quote.
Vitamin E (VE) Colorimetric Assay Kit
SKU # | E-BC-K033-S |
Detection Instrument | Spectrophotometer(533 nm) |
Detection Method | Colorimetric method |
Product Details
Properties
Synonyms | VE |
Sample Type | Serum, plasma, tissue |
Sensitivity | 0.09 μg/mL |
Detection Range | 0.09-40 μg/mL |
Detection Method | Colorimetric method |
Assay type | Quantitative |
Assay time | 60 min |
Precision | Average inter-assay CV: 4.900% | Average intra-assay CV: 4.400% |
Other instruments required | Micropipettor, Vortex mixer, Centrifuge |
Other reagents required | Normal saline (0.9% NaCl) or PBS (0.01 M, pH 7.4), Absolute ethanol,N-heptane |
Storage | 2-8℃ |
Valid period | 12 months |
Dilution of Sample
It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (0.09-40 μg/mL).
The recommended dilution factor for different samples is as follows (for reference only):
Sample type | Dilution factor |
Human serum | 1 |
Human plasma | 1 |
10% Mouse liver tissue homogenization | 1 |
Rat serum | 1 |
Rat plasma | 1 |
10% Carrot tissue homogenization | 1 |
Note: The diluent is normal saline (0.9% NaCl) or PBS (0.01 M, pH 7.4).
Detection Principle
Fe3+ can be deoxidized to Fe2+ by VE with ferroin existing. Fe2+ can react with phenanthroline and form pink compound under certain condition. VE content can be calculated by measuring the OD value at 532 nm.
Kit Components & Storage
Item | Component | Size 1 (50 assays) | Size 2 (100 assays) | Storage |
Reagent 1 | Chromogenic Agent | Powder × 1 vial | Powder × 1 vial | 2-8℃, 12 months shading light |
Reagent 2 | Ferrum Reagent | Powder × 1 vial | Powder × 1 vial | 2-8℃, 12 months shading light |
Reagent 3 | Stop Solution | 6 mL × 1 vial | 6 mL × 1 vial | 2-8℃, 12 months |
Reagent 4 | Homogenized Medium | 50 mL × 2 vials | 50 mL × 4 vials | 2-8℃, 12 months |
Reagent 5 | 1 mg/mL VE Standard | 0.4 mL × 1 vial | 0.4 mL × 1 vial | 2-8℃, 12 months shading light |
Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.
Technical Data:
Parameter:
Intra-assay Precision
Three human serum samples were assayed in replicates of 20 to determine precision within an assay (CV = Coefficient of Variation).
Parameters | Sample 1 | Sample 2 | Sample 3 |
Mean (μg/mL) | 1.50 | 18.50 | 35.00 |
%CV | 4.7 | 4.3 | 4.2 |
Inter-assay Precision
Three human serum samples were assayed 20 times in duplicate by three operators to determine precision between assays.
Parameters | Sample 1 | Sample 2 | Sample 3 |
Mean (μg/mL) | 1.50 | 18.50 | 35.00 |
%CV | 5.2 | 4.7 | 4.8 |
Recovery
Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 98%.
Sample 1 | Sample 2 | Sample 3 | |
Expected Conc. (μg/mL) | 0.1 | 0.3 | 0.5 |
Observed Conc. (μg/mL) | 0.1 | 0.299 | 0.477 |
Recovery rate (%) | 99 | 99 | 95 |
Sensitivity
The analytical sensitivity of the assay is 0.09 μg/mL VE. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.
Standard Curve:
As the OD value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only.
Concentration (μg/mL) | 0 | 5 | 10 | 15 | 20 | 25 | 30 |
Average OD | 0.029 | 0.074 | 0.130 | 0.158 | 0.213 | 0.236 | 0.284 |
Absoluted OD | 0 | 0.045 | 0.101 | 0.129 | 0.184 | 0.207 | 0.255 |