Thank you!

Your quote has been successfully submitted!

For products requiring additional information, our team will contact you within 1 business day

Failed

There was an error submitting your quote. Please try again.

Adenosine Deaminase (ADA) Activity Assay Kit– MSE Supplies LLC

Free Shipping on MSE PRO Online Orders of $500 or More! U.S. Orders Only * Offer Excludes Hazmat Shipments *

Menu

This product has been added to the cart.

Adenosine Deaminase (ADA) Activity Assay Kit

SKU: E-BC-K197-M-500

  • £56000
  • Save £9900



Adenosine Deaminase (ADA) Activity Assay Kit

SKU # E-BC-K197-M
Detection Instrument Microplate reader (550 nm)
Detection method Colorimetric method


Product Details

Properties

Synonyms ADA
Sample Type Serum, plasma, animal tissue
Sensitivity 0.03 U/L
Detection Range 0.03-99 U/L
Detection Method Colorimetric method
Assay type Enzyme Activity
Assay time 40 min
Precision Average inter-assay CV: 6% | Average intra-assay CV: 3%
Other instruments required Micropipette, Centrifuge, Water bath/Incubator
Other reagents required Normal saline (0.9% NaCl)
Storage 2-8℃
Valid period 12 months


Dilution of Sample

It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (0.03-99 U/L).

The recommended dilution factor for different samples is as follows (for reference only):

Sample type Dilution factor
10% Mouse liver tissue homogenate 1
10% Rat kidney tissue homogenate 1
10% Rat heart tissue homogenate 2-3
10% Rat spleen tissue homogenate 2-3
Mouse serum 1
Human serum 1

 

Note: The diluent is normal saline (0.9% NaCl).

 

Detection Principle

Adenosine deaminase (ADA) can hydrolyzed the substrate adenosine to form hypoxanthine riboside, which is hydrolyzed by purine riboside phosphatase to produce hypoxanthine and phosphate ribose. Under the action of xanthine oxidase, hypoxanthine produces hydrogen peroxide, which produces red substance under the action of peroxidase, 4-aminotepyrine and color source. The red substance has the maximum absorption peak at 550 nm and the changes of absorbance is proportional to the activity of ADA.

 

Kit Components & Storage

Item Component Size 1(48 T) Size 2(96 T) Storage
Reagent 1 Working Solution 10 mL × 1 vial 20 mL × 1 vial 2-8°C, 12 months, shading light
Reagent 2 Chromogenic Agent 5 mL × 1 vial 10 mL × 1 vial 2-8°C, 12 months, shading light
Reagent 3 1 mmol/L Standard 2 mL × 1 vial 4 mL × 1 vial 2-8°C, 12 months

Microplate 96 wells 96 wells No requirement

Plate Sealer 2 pieces 2 pieces

 

Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

Technical Data:

Parameter:

Intra-assay Precision

Three human serum samples were assayed in replicates of 20 to determine precision within an assay. (CV = Coefficient of Variation)

Parameters Sample 1 Sample 2 Sample 3
Mean (U/L) 5.20 34.50 75.00
%CV 3.4 2.8 2.8


Inter-assay Precision Three human serum samples were assayed 20 times in duplicate by three operators to determine precision between assays.

Parameters Sample 1 Sample 2 Sample 3
Mean (U/L) 5.20 34.50 75.00
%CV 6.1 5.9 6.0

 

Recovery

Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 98%.

Standard 1 Standard 2 Standard 3
Expected Conc. (mmol/L) 0.25 0.56 0.74
Observed Conc. (mmol/L) 0.3 0.5 0.7
Recovery rate (%) 103 95 96

 

Sensitivity

The analytical sensitivity of the assay is 0.03 U/L. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.

 

Standard Curve

As the OD value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only.

Concentration (mmol/L) 0 0.2 0.4 0.5 0.6 0.7 0.8 1.0
OD value 0.044 0.162 0.274 0.318 0.361 0.406 0.462 0.533
0.043 0.163 0.273 0.326 0.372 0.414 0.489 0.533
Average OD 0.044 0.163 0.274 0.322 0.367 0.410 0.476 0.533
Absoluted OD 0.000 0.119 0.230 0.279 0.323 0.367 0.432 0.490