Blood Ammonia Colorimetric Assay Kit

SKU: E-BC-K145-S-100

  • $ 22195
  • Save $ 3900



Blood Ammonia Colorimetric Assay Kit

SKU # E-BC-K145-S
Detection Instrument Spectrophotometer (635 nm)
Detection method Colorimetric method


Product Details

Properties

Synonyms NH3
Sample Type Serum, plasma
Sensitivity 0.01 mmol/L
Detection Range 0.01-2.0 mmol/L
Detection Method Colorimetric method
Assay type Quantitative
Assay time 65 min
Precision Average inter-assay CV: 5% | Average intra-assay CV: 4.700%
Other instruments required Micropipettor, Vortex mixer, Incubator, Centrifuge
Other reagents required Normal saline (0.9% NaCl)
Storage 2-8℃
Valid period 12 months

 

Detection Principle

Blood protein can be precipitated with protein precipitator, and enzyme activity will be destroyed, which can prevent the formation of free ammonia in vitro. Most interfering color substances were removed at the same time, indigo was formed in non-protein filtrate by Berthelot reaction, and the color depth was proportional to the content of blood ammonia. Blood ammonia content can be determined by comparing with standard solution.

 

Kit Components & Storage

Item Component Size 1
(50 assays)
Size 2
(100 assays)
Storage
Reagent 1 Acid Reagent 35 mL × 1 vial 35 mL × 2 vials 2-8°C, 12 months
Reagent 2 Chromogenic
Agent A
60 mL ×1 vial 60 mL ×2 vials 2-8°C, 12 months, shading light
Reagent 3 Chromogenic
Agent B
60 mL ×1 vial 60 mL ×2 vials 2-8°C, 12 months, shading light
Reagent 4 7 mmol/L Ammonia Standard 2 mL × 1 vial 2 mL × 1 vial 2-8°C, 12 months
Reagent 5 Standard Diluent 50 mL × 1 vial 50 mL × 1 vial 2-8°C, 12 months

 

Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

Technical Data:

Parameter:

Intra-assay Precision

Three human serum samples were assayed in replicates of 20 to determine precision within an assay. (CV = Coefficient of Variation)

Parameters Sample 1 Sample 2 Sample 3
Mean (mmol/L) 0.35 1.05 1.60
%CV 5.0 4.6 4.5


Inter-assay Precision

Three human serum samples were assayed 20 times in duplicate by three operators
to determine precision between assays.

Parameters Sample 1 Sample 2 Sample 3
Mean (mmol/L) 0.35 1.05 1.60
%CV 4.8 5.1 5.1

 

Recovery

Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 104%.

Standard 1 Standard 2 Standard 3
Expected Conc. (mmol/L) 0.65 1.2 1.7
Observed Conc. (mmol/L) 0.7 1.2 1.8
Recovery rate (%) 105 103 104

 

Sensitivity

The analytical sensitivity of the assay is 0.01 mmol/L. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.

 

Standard Curve

As the OD value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only

Concentration (mmol/L) 0 0.2 0.4 0.8 1.0 1.5 2 0
Average OD 0.010 0.075 0.140 0.269 0.332 0.492 0.663 0.010
Absoluted OD 0 0.065 0.130 0.259 0.332 0.482 0.653 0