Glucose (GLU) Fluorometric Assay Kit– MSE Supplies LLC

Free Shipping on MSE PRO Online Orders of $500 or More! U.S. Orders Only * Offer Excludes Hazmat Shipments *

Menu

This product has been added to the cart.

Glucose (GLU) Fluorometric Assay Kit

SKU: E-BC-F037-96

  • $ 34095
  • Save $ 6000


Glucose (GLU) Fluorometric Assay Kit

SKU # E-BC-F037
Detection Instrument Fluorescence microplate reader (Ex/Em=535 nm/590 nm)
Detection Method Fluorescence method


Product Details

Properties

Synonyms GLU
Sample Type Serum, plasma, urine, saliva, milk, cell
Sensitivity 0.1 μmol/L
Detection Range 0.1-20 μmol/L
Detection Method Fluorescence method
Assay Time 25 min
Precision Inter-assay CV: 2.8% | intra-assay CV: 1.7%
Storage This product can be stored at -20°C for 12 months with shading light.
Valid period 12 months

 

Dilution Of Sample

The recommended dilution factor for different samples is as follows (for reference only):

Sample type Dilution factor
Human plasma 300-600
Human serum 300-600
Chicken serum 600-1000
Human urine 1
Human milk 400-600
Saliva 3-5

 

The diluent is buffer solution. For the dilution of other sample types, please do pretest to confirm the dilution factor.

 

Detection Principle

Glucose oxidase can catalyze the oxidation of glucose into gluconic acid and produce hydrogen peroxide. In the presence of peroxidase, hydrogen peroxide reacts with the non-fluorescent substance to form fluorescent substance. The glucose content can be calculated indirectly by measuring the fluorescence intensity at the excitation wavelength of 535 nm and the emission wavelength of 590 nm.

 

Kit Components & Storage

Item Component Size 1(48 T) Size 2(96 T) Storage
Reagent 1 Buffer Solution 50 mL ×1 vial 50 mL ×2 vials -20℃, 12 months
Reagent 2 Enzyme Reagent Powder × 1 vial Powder × 1 vial -20℃, 12 months
shading light
Reagent 3 Chromogenic Agent 0.125 mL ×1 vial 0.25 mL × 1 vial -20℃, 12 months
shading light
Reagent 4 5 mmol/L Standard 0.5 mL ×1 vial 0.5 mL × 1 vial -20℃, 12 months
Black Microplate 96 wells No requirement
Plate Sealer 2 pieces


Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

    Technical Data:

    Parameter:

    Intra-assay Precision

     

    Three human serum samples were assayed in replicates of 20 to determine precision within an assay. (CV = Coefficient of Variation)

    Parameters Sample 1 Sample 2 Sample 3 
    Mean (μmol/L) 2.50 15.00 25.00
    %CV 2.0 1.5 1.6

     

    Inter-assay Precision

    Three human serum samples were assayed 20 times in duplicate by three operators to determine precision between assays.

    Parameters Sample 1 Sample 2 Sample 3 
    Mean (μmol/L) 2.50 15.00 25.00
    %CV 3.2 2.5 2.7

     

    Recovery

    Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 99%.


    Standard 1 Standard 2 Standard 3
    Expected Conc.(μmol/L) 3.5 7 12
    Observed Conc. (μmol/L) 3.5 6.9 11.9
    Recovery rate (%) 100 98 99

     

    Sensitivity

    The analytical sensitivity of the assay is 0.1 μmol/L. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.


    Standard Curve:

    As the fluorescence value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only:

    Concentration (μmol/L) 0 2 4 6 8 10 15 20
    Fluorescence value 147 427 780 1104 1446 1639 2515 3400
    148 422 791 1117 1423 1678 2546 3429
    Average fluorescence value 148 424 786 1110 1434 1658 2530 3414
    Absoluted fluorescence value 0 276 638 962 1286 1510 2382 3266