Human IL-6(Interleukin 6) ELISA Kit
SKU: E-OSEL-H0001
Human IL-6(Interleukin 6) ELISA Kit
Detection Range | 3.13-200 pg/mL |
Sensitivity | 1.02 pg/mL |
Product Details
Properties
Assay type | Sandwich-ELISA |
Format | 96T |
Assay time | 1.5h |
Detection range | 3.13-200 pg/mL |
Sensitivity | 1.02 pg/mL |
Sample type &Sample volume | serum and plasma; 50μL |
Specificity | This kit recognizes IL-6 in samples. No significant cross-reactivity or interference between IL-6 and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are < 10%. |
Application | This ELISA kit applies to the in vitro quantitative determination of IL-6 concentrations in serum and plasma. |
Test Principle
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human IL-6. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Human IL-6 are added to the micro ELISA plate wells. Human IL-6 in samples (or standards) combines with the coated antibody and HRP linked detection antibody specific to IL-6. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The concentration of Human IL-6 in the samples is then determined by comparing the OD of the samples to the standard curve.
Kit Components and Storage
An unopened kit can be stored at 2-8℃ for 6 months. After test, the unused wells and reagents should be stored according to the table.
Item | Specifications | Storage |
---|---|---|
Micro ELISA Plate(Dismountable) | 96T: 8 wells ×12 strips 48T: 8 wells ×6 strips 24T: 8 wells ×3 strips 96T*5: 5 plates, 96T |
2-8℃, 1 months |
Reference Standard | 96T: 2 vials 48T/24T: 1 vial 96T*5: 10 vials |
2-8℃, use the reconstituted standard within 24h |
Concentrated HRP Linked Detection Ab(100×) | 96T: 1 vial, 60 μL 48T/24T: 1 vial, 30 μL 96T*5: 5 vials, 60 μL |
2-8℃(Protect from light) |
Reference Standard & Sample Diluent | 96T/48T/24T: 1 vial, 20 mL 96T*5: 5 vials, 20 mL |
2-8℃ |
HRP Linked Ab Diluent | 96T/48T/24T: 1 vial, 14 mL 96T*5: 5 vials, 14 mL |
|
Concentrated Wash Buffer(25×) | 96T/48T/24T: 1 vial, 30 mL 96T*5: 5 vials, 30 mL |
|
Substrate Reagent | 96T/48T/24T: 1 vial, 10 mL 96T*5: 5 vials, 10 mL |
2-8℃(Protect from light) |
Stop Solution | 96T/48T/24T: 1 vial, 10 mL 96T*5: 5 vials, 10 mL |
2-8℃ |
Plate Sealer | 96T/48T/24T: 5 pieces 96T*5: 25 pieces |
|
Product Description | 1 copy | |
Certificate of Analysis | 1 copy |
Technical Data:
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level IL-6 were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level IL-6 were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 9.10 | 22.29 | 90.92 | 9.12 | 21.98 | 90.28 |
Standard deviation | 0.57 | 1.13 | 3.32 | 0.53 | 1.31 | 3.25 |
CV (%) | 6.24 | 5.06 | 3.65 | 5.84 | 5.98 | 3.60 |
Recovery
The recovery of IL-6 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
---|---|---|
Serum(n=8) | 89-102 | 96 |
EDTA plasma(n=8) | 92-105 | 97 |
Target Information
Database Links | SwissProt: P05231 |
Synonyms | Interleukin 6, IL6, BSF2, HGF, HSF, IFNB2, CDF |
Research Area | Signal Cancer, Metabolism, Immunology, Neuroscience |
Assay Procedures
1. Add 50µL standard or sample to the wells, Immediately add 50µL of HRP linked Detection Ab working solution to each well. Incubate for 60 min at 37°C. | |
2. Aspirate and wash the plate for 5 times. | |
3. Add 90µL of Substrate Reagent. Incubate for about 15 min at 37°C. | |
4. Add 50µL Stop Solution. | |
5. Read the plate at 450nm immediately. Calculation of the results |