Human IL-8(Interleukin 8) ELISA Kit
SKU: E-OSEL-H0014
Human IL-8(Interleukin 8) ELISA Kit
Detection Range | 15.63-1000 pg/mL |
Sensitivity | 9.38 pg/mL |
Product Details
Properties
Assay type | Sandwich-ELISA |
Format | 96T |
Assay time | 1.5h |
Detection range | 15.63-1000 pg/mL |
Sensitivity | 9.38 pg/mL |
Sample type &Sample volume | serum and plasma; 50μL |
Specificity | This kit recognizes IL-8 in samples. No significant cross-reactivity or interference between IL-8 and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are < 10%. |
Application | This ELISA kit applies to the in vitro quantitative determination of IL-8 concentrations in serum and plasma. |
Test Principle
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human IL-8. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Human IL-8 are added to the micro ELISA plate wells. Human IL-8 in samples (or standards) combines with the coated antibody and HRP linked detection antibody specific to IL-8. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The concentration of Human IL-8 in the samples is then determined by comparing the OD of the samples to the standard curve.
Kit Components and Storage
An unopened kit can be stored at 2-8℃ for 6 months. After test, the unused wells and reagents should be stored according to the table.
Item | Specifications | Storage |
---|---|---|
Micro ELISA Plate(Dismountable) | 96T: 8 wells ×12 strips 48T: 8 wells ×6 strips 24T: 8 wells ×3 strips 96T*5: 5 plates, 96T |
2-8℃, 1 months |
Reference Standard | 96T: 2 vials 48T/24T: 1 vial 96T*5: 10 vials |
2-8℃, use the reconstituted standard within 24h |
Concentrated HRP Linked Detection Ab(100×) | 96T: 1 vial, 60 μL 48T/24T: 1 vial, 30 μL 96T*5: 5 vials, 60 μL |
2-8℃(Protect from light) |
Reference Standard & Sample Diluent | 96T/48T/24T: 1 vial, 20 mL 96T*5: 5 vials, 20 mL |
2-8℃ |
HRP Linked Ab Diluent | 96T/48T/24T: 1 vial, 14 mL 96T*5: 5 vials, 14 mL |
|
Concentrated Wash Buffer(25×) | 96T/48T/24T: 1 vial, 30 mL 96T*5: 5 vials, 30 mL |
|
Substrate Reagent | 96T/48T/24T: 1 vial, 10 mL 96T*5: 5 vials, 10 mL |
2-8℃(Protect from light) |
Stop Solution | 96T/48T/24T: 1 vial, 10 mL 96T*5: 5 vials, 10 mL |
2-8℃ |
Plate Sealer | 96T/48T/24T: 5 pieces 96T*5: 25 pieces |
|
Product Description | 1 copy | |
Certificate of Analysis | 1 copy |
Technical Data:
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level IL-8 were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level IL-8 were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 54.94 | 152.77 | 425.52 | 55.10 | 154.08 | 436.69 |
Standard deviation | 3.28 | 6.97 | 16.13 | 3.07 | 8.67 | 14.89 |
CV (%) | 5.97 | 4.56 | 3.79 | 5.58 | 5.63 | 3.41 |
Recovery
The recovery of IL-8 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
---|---|---|
Serum(n=8) | 91-103 | 97 |
EDTA plasma(n=8) | 86-97 | 92 |
Target Information
Database Links | SwissProt: P10145 |
Synonyms | IL8/CXCL8 |
Research Area | Cancer, Cardiovascular, Immunology |
Assay Procedures
1. Add 50µL standard or sample to the wells, Immediately add 50µL of HRP linked Detection Ab working solution to each well. Incubate for 60 min at 37°C. | |
2. Aspirate and wash the plate for 5 times. | |
3. Add 90µL of Substrate Reagent. Incubate for about 15 min at 37°C. | |
4. Add 50µL Stop Solution. | |
5. Read the plate at 450nm immediately. Calculation of the results |