SKU # E-BC-K043-S
Detection Instrument Spectrophotometer (530 nm)
Detection Method Colorimetric method


Product Details

Properties

Sample Type Whole blood
Sensitivity 0.14 mmol/L
Detection Range 0.14-7.0 mmol/L
Detection Method Colorimetric method
Assay type Quantitative
Assay time 60 min
Precision Average inter-assay CV: 2% | Average intra-assay CV: 1.600%
Other instruments required Micropipettor, Vortex mixer, Incubator, Centrifuge
Other reagents required Normal saline (0.9% NaCl), PBS (0.01 M, pH 7.4)
Storage 2-8℃
Valid period 12 months

 

Dilution of Sample

It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (0.14-7.0 mmol/L).

The recommended dilution factor for different samples is as follows (for reference only):

Sample type Dilution factor
Rabbit whole blood 1-2
Mouse whole blood 1-2
Rat whole blood 1-2

 

Note: The diluent is normal saline (0.9% NaCl) or PBS (0.01 M, pH 7.4).

 

Detection Principle

Using NAD+ as hydrogen acceptor, LDH catalyzes the conversion of both lactate and NAD+ into pyruvic acid and NADH respectively. 1-Methoxy-5-methyl phenazine methyl sulfate (PMS) transfers hydrogen from NADH to NBT which deoxidize into purple chromogenic substrate. Lactic acid content can be calculated by measuring the OD value at 530 nm.


Kit Components & Storage

Item Component Size 1 (50 assays) Size 2 (100 assays) Storage
Reagent 1 Protein Precipitator Powder × 1 vial Powder × 2 vials 2-8℃, 12 months
Reagent 2 Enzyme Diluent 60 mL × 1 vial 60 mL × 2 vials 2-8℃, 12 months
Reagent 3 Enzyme Stock
Solution
0.6 mL × 1 vial 1.2 mL × 1 vial 2-8℃, 12 months
Reagent 4 Chromogenic Agent 12 mL × 1 vial 24 mL × 1 vial 2-8℃, 12 months
shading light
Reagent 5 3 mmol/L Standard
Solution
1 mL ×1 vial 1 mL ×1 vial 2-8℃, 12 months
Reagent 6 Stop Solution 30 mL × 1 vial 60 mL × 1 vial 2-8℃, 12 months

 

Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

Technical Data:

Parameter:

Intra-assay Precision

Three rabbit whole blood samples were assayed in replicates of 20 to determine precision within an assay (CV = Coefficient of Variation).

Parameters Sample 1 Sample 2 Sample 3
Mean (mmol/L) 1.30 3.40 5.70
%CV 1.8 1.5 1.5


Inter-assay Precision

Three rabbit whole blood samples were assayed 20 times in duplicate by three operators to determine precision between assays.

Parameters Sample 1 Sample 2 Sample 3
Mean (mmol/L) 1.30 3.40 5.70
%CV 2.1 2.3 1.6


Recovery

Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 101%.


Sample 1 Sample 2 Sample 3
Expected Conc. (mmol/L) 2.6 4.8 6.1
Observed Conc. (mmol/L) 2.6 4.8 6.3
Recovery rate (%) 99 100 104


Sensitivity