Thank you!

Your quote has been successfully submitted!

For products requiring additional information, our team will contact you within 1 business day

Failed

There was an error submitting your quote. Please try again.

Myeloperoxidase (MPO) Activity Assay Kit– MSE Supplies LLC

Free Shipping on MSE PRO Online Orders of $500 or More! U.S. Orders Only * Offer Excludes Hazmat Shipments *

Menu

Myeloperoxidase (MPO) Activity Assay Kit

SKU: E-BC-K074-S-100

To better serve you, we would like to discuss your specific requirement.
Please Contact Us for a quote.

Myeloperoxidase (MPO) Activity Assay Kit

SKU # E-BC-K074-S
Detection Instrument Spectrophotometer (460 nm)
Detection Method Colorimetric method


Product Details

Properties

Synonyms MPO
Sample Type Serum, plasma, milk, animal tissue, cells
Sensitivity 16.95 U/L
Detection Range 16.95-3349 U/L
Detection Method Colorimetric method
Assay type Enzyme Activity
Assay time 90 min
Precision Average inter-assay CV: 9.8% | Average intra-assay CV: 4.5%
Other instruments required Vortex mixer, Micropipettor, Water bath, Incubator, Centrifuge
Storage 2-8℃
Valid period 12 months

 

Dilution of Sample

It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (4.9-196.7 U/L).

The recommended dilution factor for different samples is as follows (for reference only):

Sample type Dilution factor
Human serum 1
Human plasma 1
Rat serum 1
5% Mouse brain tissue homogenization 1
5% Mouse heart tissue homogenization 1
Human milk
1

 

Note: The diluent is reagent 2 application solution.

 

Detection Principle

Myeloperoxidase reduces hydrogen peroxide to a complex. The complex react with o-dianisidine (as hydrogen donor) to produce a yellow product which has a maximum absorption peak at 460 nm. The activity of MPO can be calculated indirectly by measuring the OD value at 460nm.

 

Kit Components & Storage

Item Component Size 1 (50 assays) Size 2 (100 assays) Storage
Reagent 1 Buffer Solution 18 mL ×1 vial 36 mL ×1 vial 2-8℃, 12 months
Reagent 2 Powder A Powder × 1 vial Powder × 2 vials 2-8℃, 12 months
Reagent 3 Powder B Powder × 1 vial Powder × 2 vials 2-8℃, 12 months
Reagent 4 Saline Solution 10 mL × 1 vial 10 mL × 2 vials 2-8℃, 12 months
Reagent 5 Clarificant 12 mL ×1 vial 24 mL × 1 vial 2-8℃, 12 months
Reagent 6 Powder C Powder × 1 vial Powder × 2 vials 2-8℃, 12 months,
shading light
Reagent 7 Substrate 0.15 mL × 1 vial 0.3 mL × 1 vial 2-8℃, 12 months
Reagent 8 Acid Reagent 3 mL × 1 vial 6 mL × 1 vial 2-8℃, 12 months

 

Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.


Technical Data:

Parameter:

Intra-assay Precision

Three human serum samples were assayed in replicates of 20 to determine precision within an assay (CV = Coefficient of Variation).

Parameters Sample 1 Sample 2 Sample 3
Mean (U/L) 52.50 567.80 1126.00
%CV 4.8 4.4 4.3


Inter-assay Precision

Three human serum samples were assayed 20 times in duplicate by three operators
to determine precision between assays.

Parameters Sample 1 Sample 2 Sample 3
Mean (U/L) 52.50 567.80 1126.00
%CV 10.2 9.4 9.8

 

Recovery

Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 104%.


Sample 1 Sample 2 Sample 3
Expected Conc. (U/L) 684.5 1354 2686.5
Observed Conc. (U/L) 711.9 1435.2 2740.2
Recovery rate (%) 104 106 102


Sensitivity