Pyruvate Dehydrogenase (PDH) Activity Assay Kit
SKU: E-BC-K650-M-96
Pyruvate Dehydrogenase (PDH) Activity Assay Kit
SKU # | E-BC-K650-M |
Detection Instrument | Microplate reader (440-460 nm, optimum wavelength: 450 nm) |
Detection method | Colorimetric method |
Product Details
Properties
Synonyms | PDH |
Sample type | Serum (plasma), urine, animal tissue, cell |
Sensitivity | 0.28 U /L |
Detection range | 0.28-76.23 U/L |
Assay type | Enzyme Activity |
Assay time | 15 min |
Precision | Average inter-assay CV: 10% | Average intra-assay CV: 2% |
Other instruments required | Centrifuge, 37℃ incubator |
Storage | -20℃ |
Valid period | 12 months |
Dilution of Sample
The recommended dilution factor for different samples is as follows (for reference only):
Sample type | Dilution factor |
10% Rat liver tissue homogenate | 2-3 |
10% Rat kidney tissue homogenate | 1 |
10% Rat heart tissue homogenate | 1 |
10% Rat brain tissue homogenate | 1 |
10% Rat lung tissue homogenate | 1 |
10% Mouse liver tissue homogenate | 2-3 |
10% Mouse kidney tissue homogenate | 1 |
10% Mouse lung tissue homogenate | 1 |
1×10^6 HL-60 cells | 1 |
Rat serum | 1 |
Rat plasma | 1 |
Human urine | 1 |
Note: The diluent is extraction solution. For the dilution of other sample types, please do pretest to confirm the dilution factor.
Detection Principle
Pyruvate dehydrogenase (PDH) catalyzed the oxidative decarylation of pyruvate into acetyl CoA, which plays an important role in the metabolism of mitochondrial respiratory chain and is associated with the aging of cancer genes. Pyruvate can be decarboxylated by PDH under the action of CoA. Meanwhile, NAD+ is reduced to NADH, Which under the action of hydrogen transmitter, transfer electrons to WST-8 to produce the yellow product. The activity of PDH can be calculated by measuring the change of absorbance value at 450 nm.
Kit Components & Storage
Item | Component | Size (96 T) | Storage |
Reagent 1 | Buffer Solution | 50 mL × 2 vials | -20℃, 12 months |
Reagent 2 | Substrate | 40 mL × 1 vial | -20℃, 12 months, shading light |
Reagent 3 | Activator Agent | Powder × 2 vials | -20℃, 12 months, shading light |
Reagent 4 | Chromogenic Agent | Powder × 2 vials | -20℃, 12 months, shading light |
Reagent 5 | 20 mmol/L Standard Solution | 6 mL×1 vial | -20℃, 12 months, shading light |
Reagent 6 | Standard | Powder × 2 vials | -20℃, 12 months, shading light |
Microplate | 96 wells | No requirement | |
Plate Sealer | 2 pieces |
Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.
Technical Data:
Parameter:
Intra-assay Precision
Three human serum samples were assayed in replicates of 20 to determine precision within an assay (CV = Coefficient of Variation).
Parameters | Sample 1 | Sample 2 | Sample 3 |
Mean (U/L) | 1.50 | 35.50 | 85.60 |
%CV | 2.4 | 2.2 | 1.4 |
Inter-assay Precision
Three human serum samples were assayed 20 times in duplicate by three operators to determine precision between assays.
Parameters | Sample 1 | Sample 2 | Sample 3 |
Mean (U/L) | 1.50 | 35.50 | 85.60 |
%CV | 9.6 | 10.2 | 10.2 |
Recovery
Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 95%.
Standard 1 | Standard 2 | Standard 3 | |
Expected Conc. (mmol/L) | 0.08 | 0.23 | 0.37 |
Observed Conc. (mmol/L) | 0.1 | 0.2 | 0.4 |
Recovery rate (%) | 94 | 96 | 95 |
Sensitivity
The analytical sensitivity of the assay is 0.28 U/L. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.
Standard Curve
As the OD value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only:
Concentration (μmol/L) | 0 | 0.05 | 0.1 | 0.2 | 0.25 | 0.3 | 0.4 | 0.5 |
Average OD | 0.058 | 0.147 | 0.241 | 0.425 | 0.519 | 0.616 | 0.808 | 0.993 |
Absoluted OD | 0.000 | 0.090 | 0.184 | 0.367 | 0.461 | 0.559 | 0.750 | 0.936 |