Thank you!

Your quote has been successfully submitted!

For products requiring additional information, our team will contact you within 1 business day

Failed

There was an error submitting your quote. Please try again.

Succinate Dehydrogenase (SDH) Activity Assay Kit– MSE Supplies LLC

Free Shipping on MSE PRO Online Orders of $500 or More! U.S. Orders Only * Offer Excludes Hazmat Shipments *

Menu

This product has been added to the cart.

Succinate Dehydrogenase (SDH) Activity Assay Kit

SKU: E-BC-K649-M-96

  • $ 35795
  • Save $ 6300



Succinate Dehydrogenase (SDH) Activity Assay Kit

SKU # E-BC-K649-M
Detection Instrument Microplate reader (590-610 nm, optimum wavelength: 600 nm)
Detection method Colorimetric method


Product Details

Properties

Synonyms SDH
Sample type Animal tissue, cell
Sensitivity 0.83 U/L
Detection range 0.83-65.42 U/L
Detection Method Colorimetric method
Assay type Enzyme Activity
Assay time 10 min
Precision Average inter-assay CV: 4% | Average intra-assay CV: 2%
Other instruments required Centrifuge
Other reagents required Normal saline (0.9% NaCl), PBS(0.01 M,pH 7.4).
Storage -20℃
Valid period 12 months


Dilution of Sample

The recommended dilution factor for different samples is as follows (for reference only):

Sample type  Dilution factor
10% Rat liver tissue homogenate 1
10% Rat heart tissue homogenate 1
10% Rat kidney tissue homogenate 1
10% Mouse liver tissue homogenate 1
10% Mouse heart tissue homogenate 1
10% Mouse kidney tissue homogenate 1

 

Note: The diluent is buffer solution B. For the dilution of other sample types, please do pretest to confirm the dilution factor

 

Detection Principle

SDH catalyzes the dehydrogenation of succinate to fumarate, with electron transport materials transferring electrons to 2, 6-dichlorophenol indiophenol (DCPIP). Then, the reduced DCPIP is reduced to the oxidized DCPIP, which has a characteristic absorption peak at 600 nm. Therefore, the activity of SDH can be quantified by measure the change OD value at 600 nm.

 

Kit Components & Storage

Item Component Size 1(48 T) Size 2(96 T) Storage
Reagent 1 Buffer Solution A 50 mL × 1 vial 50 mL × 2 vials -20℃, 12 months
Reagent 2 Buffer Solution B 15 mL × 1 vial 30 mL × 1 vial -20℃, 12 months
Reagent 3 Inhibitor 0.8 mL × 1 vial 0.8 mL × 2 vials -20℃, 12 months,
shading light
Reagent 4 Substrate A 1.2 mL × 1 vial 1.2 mL × 2 vials -20℃, 12 months,
shading light
Reagent 5 Substrate B 1.2 mL × 1 vial 1.2 mL × 2 vials -20℃, 12 months,
shading light
Reagent 6 Substrate C 0.6 mL × 1 vial 1.2 mL × 1 vial -20℃, 12 months,
shading light
Microplate 96 wells No requirement
Plate Sealer 2 pieces

 

Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

Technical Data:

Parameter:

Intra-assay Precision

Three rat heart tissue samples were assayed in replicates of 20 to determine precision within an assay (CV = Coefficient of Variation).

Parameters Sample 1 Sample 2 Sample 3
Mean (U/L) 2.60 23.70 52.00
%CV 2.5 1.8 1.7


Inter-assay Precision

Three rat heart tissue samples were assayed 20 times in duplicate by three operators to determine precision between assays.

Parameters Sample 1 Sample 2 Sample 3
Mean (U/L) 2.60 23.70 52.00
%CV 3.5 4.2 4.3


Recovery

Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 105%.


Sample 1 Sample 2 Sample 3
Expected Conc. (U/L) 15 32.5 58.5
Observed Conc. (U/L) 15.6 34.1 62.0
Recovery rate (%) 104 105 106

 

Sensitivity

The analytical sensitivity of the assay is 0.83 U/L. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.