Total Oxidant Status (TOS) Colorimetric Assay Kit

SKU: E-BC-K802-M-500

  • $ 38295
  • Save $ 6800



Total Oxidant Status (TOS) Colorimetric Assay Kit

SKU # E-BC-K802-M
Detection Instrument Microplate reader (580 nm-590 nm, optimum wavelength: 590 nm)
Detection method: Colorimetric method


Product Details

Properties

Synonyms TOS
Sample type Cell, tissue, serum and other liquid samples
Sensitivity 2.5 μmol H2O2 Equiv./L
Detection range 2.5-100 μmol H2O2 Equiv./L
Detection method Colorimetric method
Assay type Quantitative
Assay time 30 min
Precision Average inter-assay CV: 3.500% | Average intra-assay CV: 2.300%
Other instruments required Micropipettor, 37℃ incubator.
Other reagents required Normal saline (0.9% NaCl), PBS (0.01 M, pH 7.4)
Storage 2-8℃
Valid period 12 months


Images

X Zhao et al investigate the function of quercetin for reducing ethanol-induced hepatocyte pyroptosis. Total oxidants status (TOS) of human L02 cell was determined using TOS colorimetric assay kit (E-BC-K802-M).

Different subscript letters indicate significant differences among the groups. (P<0.05)


Dilution of Sample

It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (2.5-100 μmol H2O2 Equiv./L).

The recommended dilution factor for different samples is as follows (for reference only):

Sample type Dilution factor
Human serum 1
Porcine serum 1
Horse serum 1
Cynomolgus macaques serum 1
10% Rat liver tissue homogenate 1
10% Rat kidney tissue homogenate 1
10% Rat lung tissue homogenate 1
10% Rat spleen tissue homogenate 1
0.612×10^6 Molt-4 cells
1

 

Note: The diluent is normal saline (0.9% NaCl) or PBS (0.01 M, pH 7.4).

 

Detection Principle

Under acid conditions, the oxidizing material in the sample can oxidize Fe2+ to Fe3+, which binds highly with xylenol orange to produce a blue-purple complex. When the pH of solution is in the range of 2-3, its maximum absorption wavelength is around 590 nm, and the color depth is proportional to the content of oxidation substances in a certain concentration and a certain time, so as to indirectly calculate the total oxidation state of the sample.

 

Kit Components & Storage

Item Component Size 1(48 T) Size 2(96 T) Storage
Reagent 1 Chromogenic Agent 12 mL × 1 vial 24 mL × 1 vial 2-8℃, 12 months,
shading light
Reagent 2 Substrate 3 mL × 1 vial 6 mL × 1 vial 2-8℃, 12 months,
shading light
Reagent 3 200 μmol/L H2O2
Standard
1 mL × 1 vial 1 mL × 1 vial 2-8℃, 12 months,
shading light
Microplate 96 wells No requirement
Plate Sealer 2 pieces


Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.


Technical Data:

Parameter:

Intra-assay Precision

Three human serum samples were assayed in replicates of 20 to determine precision within an assay (CV = Coefficient of Variation).

Parameters Sample 1 Sample 2 Sample 3
Mean (μmol H2O2 Equiv./L) 5.60 34.80 85.00
%CV 2.6 2.2 2.1


Inter-assay Precision

Three human serum samples were assayed 20 times in duplicate by three operators to determine precision between assays.

Parameters Sample 1 Sample 2 Sample 3
Mean (μmol H2O2 Equiv./L) 5.60 34.80 85.00
%CV 3.2 3.6 3.7


Sensitivity

The analytical sensitivity of the assay is 2.5 μmol H2O2 Equiv./L. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.

 

Standard Curve

As the OD value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only:

Concentration (μmol/L) 0 10 20 40 50 60 80 100
Average OD 0.085 0.146 0.195 0.337 0.439 0.502 0.668 0.767
Absoluted OD 0.000 0.061 0.110 0.252 0.354 0.417 0.584 0.682