Alkaline Phosphatase (ALP) Activity Assay Kit

SKU: E-BC-K091-M-500

  • £24700
  • Save £4400


Alkaline Phosphatase (ALP) Activity Assay Kit

SKU # E-BC-K091-M
Detection Instrument Microplate reader (500-530 nm, optimum wavelength: 520 nm)
Detection Method Colorimetric method


Product Details

Properties

Synonyms ALP
Sample Type Serum, plasma, tissue, cells
Sensitivity 0.13 King unit/100 mL
Detection Range 0.13-50 King unit/100 mL
Detection Method Colorimetric method
Assay type Enzyme Activity
Assay time 30 min
Precision Average inter-assay CV: 8.500% | Average intra-assay CV: 5.100%
Other instruments required Micropipettor, Centrifuge, Incubator, Vortex mixer
Other reagents required Normal saline (0.9% NaCl), PBS (0.01 M, pH 7.4)
Storage 2-8℃
Valid period 12 months

 

Dilution of Sample

It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (0.13-50 King unit/100 mL).

The recommended dilution factor for different samples is as follows (for reference only):

Sample type Dilution factor
Human plasma 1
Human urine 1
Rat serum 1
Cells culture supernatant 1
10% Mouse kidney tissue homogenate 30-50
0% Mouse liver tissue homogenate 1
10% Mouse brain tissue homogenate 1
HepG2 cells 1

 

Note: The diluent is normal saline (0.9% NaCl), PBS (0.01 M, pH 7.4).

 

Detection Principle

Alkaline phosphatase decompose benzene disodium phosphate to produce free phenol and phosphoric acid. Phenol react with 4-aminopyrline in alkaline solution and oxidizes with potassium ferricyanide to form red quinone derivative. The enzyme activity can be calculated indirectly by measuring the OD value.

 

Kit Components & Storage

Item Component Size 1(96 T) Size 2 (500 Assays) Storage
Reagent 1 Buffer Solution 3 mL ×1 vial 15 mL ×1 vial 2-8℃, 12 months,
shading light
Reagent 2 Substrate Solution 3 mL × 1 vial 15 mL ×1 vial 2-8℃, 12 months,
shading light
Reagent 3 Chromogenic Agent 18 mL × 1 vial 50 mL × 2 vials 2-8℃, 12 months,
shading light
Reagent 4 0.5 mg/mL Phenol Standard 1.5 mL × 1 vial 5 mL × 1 vial 2-8℃, 12 months,
shading light

Microplate 96 wells
No requirement

Plate Sealer 2 pieces

 

Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

Technical Data:

Parameter:

Intra-assay Precision

Three human serum samples were assayed in replicates of 20 to determine precision within an assay (CV = Coefficient of Variation).

Parameters Sample 1 Sample 2 Sample 3
Mean (King unit/100 mL) 1.40 24.60 41.50
%CV 5.4 4.9 5.0


Inter-assay Precision

Three human serum samples were assayed 20 times in duplicate by three operators to determine precision between assays.

Parameters Sample 1 Sample 2 Sample 3
Mean (King unit/100 mL) 1.40 24.60 41.50
%CV 8.7 8.0 8.8

 

Recovery

Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 94%.


Sample 1 Sample 2 Sample 3
Expected Conc. (mg/mL) 0.036 0.17 0.38
Observed Conc. (mg/mL) 0.0 0.2 0.4
Recovery rate (%) 95 93 94


Sensitivity

The analytical sensitivity of the assay is 0.13 King unit/100 mL. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.

 

Standard Curve

As the OD value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only

Concentration (mg/mL) 0 0.025 0.05 0.1 0.2 0.3 0.4 0.5
Average OD 0.054 0.135 0.215 0.371 0.664 0.965 1.265 1.563
Absoluted OD 0 0.081 0.161 0.317 0.610 0.911 1.211 1.509