Thank you!

Your quote has been successfully submitted!

For products requiring additional information, our team will contact you within 1 business day

Failed

There was an error submitting your quote. Please try again.

Aspartate Aminotransferase (AST/GOT) Activity Fluorometric Assay Kit– MSE Supplies LLC

Free Shipping on MSE PRO Online Orders of $500 or More! U.S. Orders Only * Offer Excludes Hazmat Shipments *

Menu

This product has been added to the cart.

Aspartate Aminotransferase (AST/GOT) Activity Fluorometric Assay Kit

SKU: E-BC-F043-96

  • £26000
  • Save £4600


Aspartate Aminotransferase (AST/GOT) Activity Fluorometric Assay Kit

SKU # E-BC-F043
Detection Instrument Fluorescence microplate reader (Ex/Em=535 nm/587 nm)
Detection Method Fluorescence method


Product Details

Properties

Synonyms AST;GOT
Sample Type Cell, animal tissue, serum (plasma), other liquid samples
Sensitivity 0.03 U/L
Detection Range 0.03-1.5 U/L
Detection Method Fluorescence method
Assay Time 80 min
Precision Inter-assay CV: 7.3% | intra-assay CV: 2.2%
Storage This product can be stored at -20°C for 12 months with shading light.
Valid period 12 months

 

Dilution Of Sample

The recommended dilution factor for different samples is as follows (for reference only):

Sample type Dilution factor
10% Rat heart tissue homogenate 300-600
10% Rat brain tissue homogenate 300-500
10% Rat liver tissue homogenate 200-500
Human serum 5-15
10% Rat kidney tissue homogenate 200-500
Rat serum 10-20
10% Rat spleen tissue homogenate 50-100
293T cell(1×10^6) 10-20
10% Mouse lung tissue homogenate 100-200
293T cell supernatant 5-15

 

The diluent is buffer solution. For the dilution of other sample types, please do pretest to confirm the dilution factor.

 

Detection Principle

AST can catalyze the substrate to produce pyruvate. Pyruvate is oxidized to produce hydrogen peroxide, which causes the fluorescent probe to produce fluoresce. The activity of AST can be calculated by measuring the increase of fluorescence value at the excitation wavelength of 535 nm and the emission wavelength of 587 nm.

 

Kit Components & Storage

Item Component Size 1(48 T) Size 2(96 T) Storage
Reagent 1 Buffer Solution 60 mL × 1 vial 60 mL × 2 vials -20℃, 12 months
Reagent 2 Substrate 1.5 mL × 1 vial 1.5 mL × 2 vials -20℃, 12 months shading light
Reagent 3 Enzyme Reagent Powder × 1 vial Powder × 2 vials -20℃, 12 months shading light
Reagent 4 Aqueous Alkali 0.3 mL × 1 vial 0.6 mL × 1 via -20℃, 12 months shading light
Reagent 5 100 mmol/L Pyruvate Standard 1 mL × 1 vial 1 mL × 1 vial -20℃, 12 months shading light
Black Microplate 96 wells No requirement
Plate Sealer 2 pieces


Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

    Technical Data:

    Parameter:

    Intra-assay Precision

    Three human serum samples were assayed in replicates of 20 to determine precision within an assay. (CV = Coefficient of Variation)

    Parameters Sample 1 Sample 2 Sample 3
    Mean (U/L) 0.10 0.80 1.20
    %CV 2.5 2.2 1.9

     

    Inter-assay Precision

     

    Three human serum samples were assayed 20 times in duplicate by three operators to determine precision between assays.

    Parameters Sample 1 Sample 2 Sample 3
    Mean (U/L) 0.10 0.80 1.20
    %CV 6.7 7.5 7.7


    Recovery

    Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 100%.


    Standard 1 Standard 2 Standard 3
    Expected Conc. (μmol/L) 15 36.5 55
    Observed Conc. (μmol/L) 14.9 35.8 56.7
    Recovery rate (%) 99 98 103

     

    Sensitivity

    The analytical sensitivity of the assay is 0.03 U/L. This was determined by adding two standard deviations to the mean F. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.


    Standard Curve:

    As the fluorescence value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only:

    Concentration (μmol/L) 0 10 20 30 40 50 60 80
    Average fluorescence value 1019 2598 4185 5490 7018 8482 9917 12725
    Absoluted fluorescence value 0 1579 3166 4471 5999 7463 8898 11706