BCA Protein Colorimetric Assay Kit– MSE Supplies LLC

Free Shipping on MSE PRO Online Orders of $500 or More! U.S. Orders Only * Offer Excludes Hazmat Shipments *

Menu

This product has been added to the cart.

BCA Protein Colorimetric Assay Kit

SKU: E-BC-K318-M-500

  • £10400
  • Save £1900



BCA Protein Colorimetric Assay Kit

SKU # E-BC-K318-M
Detection Instrument Microplate reader (540-590 nm, optimum wavelength: 562 nm)
Detection method Colorimetric method


Product Details

Properties

Synonyms TP
Sample type Serum, plasma, cell culture supernatant, tissue, cells
Sensitivity 0.0165 mg/mL
Detection range 0.0165-1 mg/mL
Detection Method Colorimetric method
Assay type Quantitative
Assay time 40 min
Precision Average inter-assay CV: 4.500% | Average intra-assay CV: 2.200%
Other instruments required Vortex mixer, Micropipettor, Incubator
Other reagents required Normal saline (0.9% NaCl), PBS (0.01 M, pH 7.4)
Storage RT
Valid period 12 months


Dilution of Sample

It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (0.0165-1 mg/mL).

The recommended dilution factor for different samples is as follows (for reference only):

Sample type Dilution factor
10% Mouse brain tissue homogenization 8-12
10% Mouse kidney tissue homogenization 8-12
Human serum 100-200
10% Rat liver tissue homogenization 15-20
10% Mouse heart tissue homogenization 8-12
Rat serum 100-200

 

Note: The diluent is normal saline (0.9% NaCl) or PBS (0.01 M, pH 7.4).

 

Detection Principle

Cu2+ can be reduced to Cu+ by protein in alkaline condition. Cu+ can combine with BCA reagent and form purple complex, which has a maximum absorption peak at 562 nm. The absorbance value is proportional to the protein concentration. Therefore, the protein concentration can be calculated according to the OD value.


Kit Components & Storage

Item Component Size 1 (96 T) Size 2
(500 Assays)
Storage
Reagent 1 BCA Reagent 25 mL × 1 vial 50 mL × 2 vials RT, 12 months
Reagent 2 Copper Salt Solution 0.5 mL × 1 vial 3 mL × 1 vial RT, 12 months
Reagent 3 Standard 1 mg × 1 vial 1 mg × 5 vials RT, 12 months
Reagent 4 Standard Diluent 15 mL × 1 vial 30 mL × 1 vial RT, 12 months

Microplate 96 wells
No requirement

Plate Sealer 2 pieces

 

Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

Technical Data:

Parameter:

Intra-assay Precision

Three human serum samples were assayed in replicates of 20 to determine precision within an assay. (CV = Coefficient of Variation)

Parameters Sample 1 Sample 2 Sample 3
Mean (mg/mL) 0.08 0.28 0.65
%CV 2.5 2.3 1.8


Inter-assay Precision

Three human serum samples were assayed 17 times in duplicate by three operators to determine precision between assays.

Parameters Sample 1 Sample 2 Sample 3
Mean (mg/mL) 0.08 0.28 0.65
%CV 4.2 4.5 4.8


Recovery

Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 100%.

Parameters Standard 1 Standard 2 Standard 3
Expected Conc. (mg/mL) 0.25 0.45 0.85
Observed Conc. (mg/mL) 0.3 0.4 0.9
Recovery rate (%) 101 99 100


Sensitivity

The analytical sensitivity of the assay is 0.0165 mg/mL. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.


Standard Curve

As the OD value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only:

Concentration
(mg/mL)
0 0.2 0.3 0.4 0.6 0.7 0.9 1
Average OD 0.081 0.308 0.395 0.494 0.666 0.761 0.906 0.984
Absoluted OD 0 0.227 0.314 0.413 0.585 0.680 0.825 0.903