Blood Ammonia Colorimetric Assay Kit
SKU: E-BC-K145-M-96
Blood Ammonia Colorimetric Assay Kit
SKU # | E-BC-K145-M |
Detection Instrument | Microplate reader (600-660 nm, optimum wavelength: 635 nm) |
Detection method | Colorimetric method |
Product Details
Properties
Synonyms | NH3 |
Sample Type | Serum, plasma |
Sensitivity | 0.01 mmol/L |
Detection Range | 0.01-2.5 mmol/L |
Detection Method | Colorimetric method |
Assay type | Quantitative |
Assay time | 45 min |
Precision | Average inter-assay CV: 7.200% | Average intra-assay CV: 4.100% |
Other instruments required | Micropipettor, Centrifuge, Incubator, Vortex mixer |
Other reagents required | Normal saline (0.9% NaCl) |
Storage | 2-8℃ |
Valid period | 12 months |
Dilution of Sample
It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (0.01-2.5 mmol/L).
The recommended dilution factor for different samples is as follows (for reference only):
Sample type | Dilution factor |
Human serum | 1 |
Human plasma | 1 |
Mouse serum | 1 |
Rat plasma | 1 |
Dog serum | 1 |
Horse serum | 1 |
Note: The diluent is double distilled water or normal saline (0.9% NaCl).
Detection Principle
Blood protein can be precipitated with protein precipitator, and enzyme activity will be destroyed, which can prevent the formation of free ammonia in vitro. Most interfering color substances were removed at the same time, indigo was formed in non-protein filtrate by Berthelot reaction, and the color depth was proportional to the content of blood ammonia. Blood ammonia content can be determined by comparing with standard solution.
Kit Components & Storage
Item | Component | Size 1(48 T) | Size 2(96 T) | Storage |
Reagent 1 | Acid Reagent | 20 mL × 1 vial | 40 mL × 1 vial | 2-8°C, 12 months |
Reagent 2 | Chromogenic Agent A |
10 mL ×1 vial | 20 mL ×1 vial | 2-8°C, 12 months, shading light |
Reagent 3 | Chromogenic Agent B |
10 mL ×1 vial | 20 mL ×1 vial | 2-8°C, 12 months, shading light |
Reagent 4 | 7 mmol/L Standard | 1.5 mL × 1 vial | 1.5 mL × 1 vial | 2-8°C, 12 months |
Reagent 5 | Standard Diluent | 30 mL × 1 vial | 30 mL × 1 vial | 2-8°C, 12 months |
Microplate | 96 wells | No requirement | ||
Plate Sealer | 2 pieces |
Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.
Technical Data:
Parameter:
Intra-assay Precision
Three human serum samples were assayed in replicates of 20 to determine precision within an assay. (CV = Coefficient of Variation)
Parameters | Sample 1 | Sample 2 | Sample 3 |
Mean (mmol/L) | 0.45 | 1.26 | 2.00 |
%CV | 4.2 | 4.3 | 3.8 |
Inter-assay Precision
Three human serum samples were assayed 20 times in duplicate by three operators to determine precision between assays.
Parameters | Sample 1 | Sample 2 | Sample 3 |
Mean (mmol/L) | 0.45 | 1.26 | 2.00 |
%CV | 7.1 | 7.5 | 7.0 |
Recovery
Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 100%.
Standard 1 | Standard 2 | Standard 3 | |
Expected Conc. (mmol/L) | 0.35 | 0.88 | 1.7 |
Observed Conc. (mmol/L) | 0.4 | 0.9 | 1.8 |
Recovery rate (%) | 101 | 104 | 104 |
Sensitivity
The analytical sensitivity of the assay is 0.01 mmol/L. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.
Standard Curve
As the OD value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only:
Concentration (mmol/L) | 0 | 0.2 | 0.4 | 0.8 | 1.0 | 1.5 | 2 | 2.5 |
Average OD | 0.054 | 0.135 | 0.215 | 0.361 | 0.413 | 0.601 | 0.767 | 0.965 |
Absoluted OD | 0 | 0.081 | 0.161 | 0.307 | 0.359 | 0.547 | 0.713 | 0.911 |