Thank you!

Your quote has been successfully submitted!

For products requiring additional information, our team will contact you within 1 business day

Failed

There was an error submitting your quote. Please try again.

Cell Cycle Assay Kit (Green Fluorescence)– MSE Supplies LLC

Free Shipping on MSE PRO Online Orders of $500 or More! U.S. Orders Only * Offer Excludes Hazmat Shipments *

Menu

Cell Cycle Assay Kit (Green Fluorescence)

SKU: E-CK-A352-100

To better serve you, we would like to discuss your specific requirement.
Please Contact Us for a quote.

Cell Cycle Assay Kit (Green Fluorescence)

 

Introduction

Elabscience® Cell Cycle Assay Kit (Fluorometric-Red) is a kit that detects cell cycle by detecting DNA content. Cell cycle refers to the whole process from the end of one mitosis to the end of the next. During this process, the genetic material is replicated and doubled, and evenly distributed to two daughter cells at the end of division. Cell cycle can be divided into phases like interphase and Metaphase. Intercellular phase can also be divided into dormancy (zero gap, G0), prophase of DNA synthesis (first gap, G1), anaphase of DNA synthesis (synthesis, S) and anaphase of DNA synthesis (second gap, G2). DNA can bind to some specific fluorescent dyes (such as Cycle Green), the fluorescent dyes binding to DNA at different stages are different, and the fluorescence intensity detected by flow cytometry can also be used to detect different phases in cell cycle.

After staining with Cycle Green, assuming that the fluorescence intensity of G0/G1 phase cells is 1, the theoretical value of fluorescence intensity of G2/M phase cells containing double genomic DNA is 2, and the fluorescence intensity of S phase cells undergoing DNA replication is between 1 and 2. Apoptotic cells lost part of genomic DNA fragmentation due to nucleus concentration and DNA fragmentation. Therefore, apoptotic cells showed obvious weak staining after Cycle Green staining and the fluorescence intensity was less than 1. The sub-G1 peak appeared on the flow cytometry result which is apoptotic cell. The kit can be used to detect the DNA content (cell cycle) of cultured cells (suspension or adherence cells).

Jurkat cells were treated with 70% ethanol overnight and detected with this kit.

 

Product Details

Properties

Application Cell Cycle
Detection method Fluorometric Method
Sample type Cell samples
Assay time 3 hours
Detection instrument Flow Cytometer
Dye Type Cycle Green
Ex/Em (nm) 500/530
Channel set FITC
Other reagents required Ethanol, PBS(E-IR-R187)
Storage -20°C, shading light.
Shipping Ice bag
Expiration date 12 months

 

Staining Procedure

  1. Reagent Preparation
    A. Store the absolute ethanol at -20°C overnight.
    B. Take out the RNase A reagent dissolve fully, mix it and put on ice for use.
  2. Sample Preparation
    1) Collect 5×105 cells for each, centrifuge at 300×g for 5 min and discard the supernatant.
    2) Add 1 mL PBS to resuspend gently. Centrifuge at 300×g for 5 min and discard the supernatant.
    3) Add 0.3 mL PBS to resuspend the cells.
  3.  Add 0.7 mL absolute ethanol from -20°C refrigerator, mix fully and store at -20°C for 1 h or overnight.
  4. Centrifuge at 300×g for 5 min and discard the supernatant. Add 1 mL PBS to resuspend the cells, store at RT for 15 min.
  5. Centrifuge at 300×g for 5 min and discard the supernatant. Add 100 μL RNase A reagent to resuspend the cells, incubate at 37°C water bath for 30 min.
  6. Add 400 μL Cycle Green Reagent, mix fully and incubate at 2-8°C for 30 min in the dark.
  7. Analyze the cells immediately with proper machine settings. The maximum excitation wavelength of Cycle Green is 500 nm, and the maximum emission wavelength is 530 nm.

 

Storage

RNase A Reagent should be stored at -20°C for 1 year. Cycle Green Reagent should be stored at -20°C in dark for 1 year.

 

Notes

  1. For maximal assay performance, this kit should be used within 12 months.
  2. The experimental results need to be detected by flow cytometer.
  3. Avoid extended exposure of the samples to direct light to protect the fluorophores from quenching.
  4. Cycle Green Reagent stimulates the human body. Please be careful during operation and pay attention to proper protection to avoid direct contact with the human body or inhalation.
  5. This kit is for research use only. For your safety and health, please wear lab clothes and gloves. Instructions should be followed strictly, changes of operation may result in unreliable results.