Xanthine Oxidase (XOD) Activity Fluorometric Assay Kit
SKU: E-BC-F019-500
Xanthine Oxidase (XOD) Activity Fluorometric Assay Kit
SKU # | E-BC-F019 |
Detection Instrument | Fluorescence microplate reader (Ex/Em=535 nm/587 nm) |
Detection Method | Fluorescence method |
Product Details
Properties
Synonyms | XOD |
Sample Type | Serum, plasma, animal tissue |
Sensitivity | 0.01 U/L |
Detection Range | 0.01 -1.2 U/L |
Detection Method | Fluorescence method |
Assay Time | 25 min |
Precision | Inter-assay CV: 9% | intra-assay CV: 4.1% |
Storage | This product can be stored at -20°C for 12 months with shading light. |
Valid period | 12 months |
Dilution Of Sample
The recommended dilution factor for different samples is as follows (for reference only):
Sample type | Dilution factor |
---|---|
Mouse serum | 1 |
Dog serum | 3-5 |
Rat plasma | 1 |
Horse serum | 1 |
Human plasma | 1 |
10% Rat kidney tissue homogenate | 5-10 |
10% Mouse heart tissue homogenate | 5-10 |
10% Rat lung tissue homogenate | 5-10 |
The diluent is buffer solution. For the dilution of other sample types, please do pretest to confirm the dilution factor.
Detection Principle
Hypoxanthine are oxidized by xanthine oxidase (XOD) to produce xanthine and super oxygen anion, which will quickly convert to hydrogen peroxide in the system, and then, in the role of peroxidase, hydrogen peroxide can oxidize the non-fluorescent probe to fluorescent substance. By measuring the fluorescence value, the corresponding the activity of xanthine oxidase can be calculated.
Kit Components & Storage
Item | Component | Size 1 (96 T) | Size 2 (500 Assays) | Storage |
Reagent 1 | Buffer Solution | 60 mL × 1 vial | 60 mL × 5 vials | -20℃, 12 months |
Reagent 2 | Probe Solution | 0.3 mL × 1 vial | 1.5 mL × 1 vial | 2-8°C, 12 months shading light |
Reagent 3 | Enzyme Reagent 1 | Powder ×1 vial | Powder ×5 vials | -20℃, 12 months |
Reagent 4 | Substrate | 4 mL × 1 vial | 20 mL × 1 vial | 2-8°C, 12 months shading light |
Reagent 5 | 2 mmol/L H2O2 Standard Solution | 1.5 mL × 1 vial | 7 mL × 1 vial | -20℃, 12 months |
Black Microplate | 96 wells | No requirement | ||
Plate Sealer | 2 pieces |
Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.
Technical Data:
Parameter:
Intra-assay Precision
Three human serum samples were assayed in replicates of 20 to determine precision within an assay. (CV = Coefficient of Variation)
Parameters | Sample 1 | Sample 2 | Sample 3 |
Mean (U/L) | 0.10 | 0.50 | 1.00 |
%CV | 4.8 | 4.2 | 3.3 |
Inter-assay Precision
Three human serum samples were assayed 20 times in duplicate by three operators to determine precision between assays.
Parameters | Sample 1 | Sample 2 | Sample 3 |
Mean (U/L) | 0.10 | 0.50 | 1.00 |
%CV | 8.5 | 10.0 | 10.0 |
Sensitivity
The analytical sensitivity of the assay is 0.01 U/L. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.
Standard Curve:
As the fluorescence value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only
Concentration (μmol/L) | 0 | 1 | 2 | 4 | 6 | 8 | 10 | 12 |
Fluorescence value | 156 | 616 | 1161 | 2311 | 3506 | 4722 | 5740 | 7092 |
157 | 634 | 1224 | 2443 | 3581 | 4693 | 5874 | 6873 | |
Average fluorescence value | 156 | 625 | 1193 | 2377 | 3543 | 4707 | 5807 | 6982 |
Absoluted fluorescence value | 0 | 469 | 1036 | 2221 | 3387 | 4551 | 5651 | 6826 |