SKU: E-BC-K083-M-96
α-Ketoglutarate Dehydrogenase(α-KGDH) Activity Assay Kit
SKU # | E-BC-K083-M |
Detection Instrument | Microplate reader (440-460 nm, optimum wavelength: 450 nm) |
Detection Method | Colorimetric method |
Product Details
Properties
Synonyms | α-KGDH |
Sample Type | Serum, plasma, animal and plant tissue |
Sensitivity | 8.3 U/L |
Detection Range | 8.3-42.3 U/L |
Detection Method | Colorimetric method |
Assay type | Enzyme Activity |
Assay time | 40 min |
Precision | Average inter-assay CV: 8.500% | Average intra-assay CV: 1.500% |
Other instruments required | Incubator, Centrifuge |
Storage | 2-8℃ |
Valid period | 12 months |
Detection Principle
α-ketoglutarate dehydrogenase (α-KGDH) is a key enzyme in the tricarboxylic acid (TCA) cycle. While the substrate being catalyzed by α-KGDH, NAD+ is reduced to NADH, and electrons are transferred to WST-8 under the action of hydrogen transmitter to produce yellow product. The activity of α-KGDH can be calculated by measuring the change of absorbance value at 450 nm.
Kit Components & Storage
Item | Component | Size 1(96 T) | Storage |
Reagent 1 | Extraction Solution | 55 mL ×1 vial | -20℃, 12 months |
Reagent 2 | Buffer Solution | 28 mL × 1 vial | -20℃, 12 months |
Reagent 3 | Substrate | Powder × 1 vial | -20 ℃, 12 months, shading light |
Reagent 4 | Chromogenic Agent | 3 mL × 1 vial | -20 ℃, 12 months, shading light |
Reagent 5 | Clarificant | 3 mL × 1 vial | -20℃, 12 months |
Reagent 6 | Standard | Powder × 1 vial | -20℃, 12 months |
Microplate | 96 wells | No requirement | |
Plate Sealer | 2 pieces |
Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.
Technical Data:
Parameter:
Intra-assay Precision
Three human serum samples were assayed in replicates of 20 to determine precision within an assay (CV = Coefficient of Variation).
Parameters | Sample 1 | Sample 2 | Sample 3 |
Mean (U/L) | 12.50 | 27.60 | 33.80 |
%CV | 1.7 | 1.4 | 1.4 |
Inter-assay Precision
Three human serum samples were assayed 20 times in duplicate by three operators to determine precision between assays.
Parameters | Sample 1 | Sample 2 | Sample 3 |
Mean (U/L) | 12.50 | 27.60 | 33.80 |
%CV | 8.2 | 8.5 | 8.8 |
Recovery
Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 100%.
Sample 1 | Sample 2 | Sample 3 | |
Expected Conc. (mmol/L) | 0.15 | 0.28 | 0.37 |
Observed Conc. (mmol/L) | 0.2 | 0.3 | 0.4 |
Recovery rate (%) | 101 | 98 | 101 |
Sensitivity
The analytical sensitivity of the assay is 8.3 U/L. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.
Standard Curve
As the OD value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only:
Concentration (mmol/L) | 0 | 0.1 | 0.2 | 0.25 | 0.3 | 0.35 | 0.4 | 0.5 |
Average OD | 0.091 | 0.238 | 0.436 | 0.545 | 0.601 | 0.681 | 0.811 | 1.017 |
Absoluted OD | 0 | 0.147 | 0.345 | 0.453 | 0.509 | 0.589 | 0.719 | 0.925 |