Thank you!

Your quote has been successfully submitted!

For products requiring additional information, our team will contact you within 1 business day

Failed

There was an error submitting your quote. Please try again.

Malondialdehyde (MDA) Colorimetric Assay Kit (Cell Samples)– MSE Supplies LLC

Free Shipping on MSE PRO Online Orders of $500 or More! U.S. Orders Only * Offer Excludes Hazmat Shipments *

Menu
Malondialdehyde (MDA) Colorimetric Assay Kit (Cell Samples) - MSE Supplies LLC

Malondialdehyde (MDA) Colorimetric Assay Kit (Cell Samples)

SKU: E-BC-K028-M-500

To better serve you, we would like to discuss your specific requirement.
Please Contact Us for a quote.

Malondialdehyde (MDA) Colorimetric Assay Kit (Cell Samples)

SKU # E-BC-K028-M
Detection Instrument Microplate reader (530-540 nm, optimum wavelength: 532 nm)
Detection Method Colorimetric method


Product Details

Properties

Synonyms MDA
Sample Type Cells
Sensitivity 0.29 nmol/mL
Detection Range 0.29-100 nmol/mL
Detection Method Colorimetric method
Assay type Quantitative
Assay time 65 min
Precision Average inter-assay CV: 3.500% | Average intra-assay CV: 3.300%
Other instruments required Micropipettor, Centrifuge, Incubator, Water bath, Vortex mixer
Other reagents required Absolute ethanol
Storage 2-8℃
Valid period 12 months


Dilution of Sample

It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (0.29-100 nmol/mL).

Sample type Dilution factor
A549 cells 1
HepG2 cells 1
293T cells 1

 

Note: The diluent is reagent 5.

 

Detection Principle

MDA in the catabolite of lipid peroxide can react with thiobarbituric acid (TBA) and produce red compound, which has a maximum absorption peak at 532 nm.

 

Kit Components & Storage

Item Component Size (500 Assays) Storage
Reagent 1 Clarificant 30 mL × 1 vial 2-8℃, 12 months
Reagent 2 Acid Reagent 30 mL × 1 vial 2-8℃, 12 months
Reagent 3 Chromogenic Agent 50 mL × 3 vials 2-8℃, 12 months shading light
Reagent 4 10 nmol/mL Standard 5 mL × 1 vial 2-8℃, 12 months
Reagent 5 Extracting Solution 30 mL × 1 vial 2-8℃, 12 months

 

Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

Technical Data:

Parameter:

Intra-assay Precision

Three HepG2 cell samples were assayed in replicates of 20 to determine precision within an assay (CV = Coefficient of Variation).

Parameters Sample 1 Sample 2 Sample 3
Mean (nmol/mL) 2.60 34.50 88.50
%CV 3.6 3.2 3.1


Inter-assay Precision

Three HepG2 cell samples were assayed 20 times in duplicate by three operators to determine precision between assays.

Parameters Sample 1 Sample 2 Sample 3
Mean (nmol/mL) 2.60 34.50 88.50
%CV 3.1 3.8 3.6


Recovery

Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 95%.


Sample 1 Sample 2 Sample 3
Expected Conc. (nmol/mL) 12 49 91
Observed Conc. (nmol/mL) 11.2 48.0 85.5
Recovery rate (%) 93 98 94


Sensitivity

The analytical sensitivity of the assay is 0.29 nmol/mL MDA. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.