MSE PRO 190-900nm Atomic Absorption Spectrophotometer
SKU: BB1835
MSE PRO™ 190-900nm Atomic Absorption Spectrophotometer
The MSE PRO™ Atomic Absorption Spectrophotometer analytical instrument is a PC-controlled system designed for high-precision flame and graphite furnace atomic absorption spectroscopy. It features advanced automation, including automatic wavelength scanning, peak searching, and position adjustments for optimized burner performance. With integrated safety chains protection, it ensures secure operation against gas leakage, under-voltage, and abnormal flameout. The system supports flame absorption and emission methods, making it highly versatile. Data processing capabilities allow for the storage of over 500 results, with Excel-format output for easy analysis and reporting.
- Fully Automated & PC-Controlled: Offers automatic wavelength scanning, peak searching, and burner position adjustments for seamless operation.
- High-Stability Optical System: Integrated floated optical platform design enhances shock resistance and measurement accuracy.
- Advanced Safety Features: Includes automatic warnings and protection against gas leakage, incorrect burner connections, and flameout for a secure working environment.
- Multi-Method Measurement: Supports both flame absorption and emission methods, making it ideal for diverse analytical needs.
- Efficient Data Management: Stores over 500 analysis results with Excel-format export, enabling easy result tracking and report generation.
Brand: MSE PRO™
Manufacturer: MSE Supplies LLC
MSE Supplies offers a variety of Atomic Absorption Spectrophotometers. Please contact us for more information.
Specifications:
SKU# | BB1835 |
Wavelength Range | 190-900nm |
Spectral Bandwidth | 0.1, 0.2, 0.4, 1.0 and 2.0nm |
Wavelength Accuracy | ≤±0.15nm |
Wavelength Repeatability | ≤0.04nm |
Base Line Stability | ≤0.002A/30min (Cu) |
Characteristic Concentration | 0.02μg/ml/1% (Cu) |
Detection Limit | 0.004μg/ml (Cu) |
Preciseness | ≤0.5% |
Grating | 1800 lines/mm |
Inflamer | All-metal titanium burner |
Atomizer | Effective glass atomizer |
Lamp Stand | 8 |
D2 background Calibration | When the background is 1 A, the background ability should be deducted not less than 50 times; self-absorption background deduction method |
Power Supply | 220V |
Standard Accessories | Software; Inkjet printer; Oil free Air compressor; Acetylene reducing valve; Cu Hollow cathode lamp; Air filter |
Optional Accessories | Graphite Furnace; Hollow cathode lamp (Ag, Au, Ca, Cd, Cr, Fe, Hg, K, Au, Mg, Mn, Na, Ni, Pb, Zn, Sb, etc), Graphite tubes, Recirculating cooling water system, Hydride generator, Autosampler |
Flame System:
Acetylene Air Burner | 100mm |
Ignition Dynamic Baseline Drift | ≤0.006A/30min |
Characteristic Concentration | ≤0.025μg/ml/1%(Cu) |
Related Standard Deviation of the Accuracy | ≤0.5%(Cu, absorbance>0.8A) (detection limit Cu≤0.008μg/ml) |
Safety System | Can automatically cut off the gas when the pressure is not enough, the power is off, flameout and unconformity of the burner. |
Graphite Furnace (Optional):
Max. Temperature | 3000°C |
Max. Heating Speed | ≥2000°C/S |
Characteristic Quantity | Cd≤0.5*10-12g Cu≤0.5*10-11g |
Accuracy | Cu≤3% Cd≤3% |
Safety System | Over current protection Low air pressure alarm/protection; Low cooling water flow alarm/protection |
Power Supply | 220V |
Consumption | 7000W |