Urea (BUN) Colorimetric Assay Kit (Urease Method)

SKU: E-BC-K183-S-100

To better serve you, we would like to discuss your specific requirement.
Please Contact Us for a quote.

Urea (BUN) Colorimetric Assay Kit (Urease Method)

SKU # E-BC-K183-S
Detection Instrument Spectrophotometer (580 nm)
Detection method Colorimetric method


Product Details

Properties

Synonyms BUN
Sample type Serum, plasma, urine, saliva, milk
Sensitivity 0.114 mmol/L
Detection range 0.114-30 mmol/L
Detection method Colorimetric method
Assay type Quantitative
Assay time 60 min
Precision Average inter-assay CV: 4.700% | Average intra-assay CV: 4.600%
Other instruments required Micropipettor, Vortex mixer, Incubator
Other reagents required Normal saline (0.9% NaCl), PBS (0.01 M, pH 7.4)
Storage 2-8℃
Valid period 12 months


Dilution of Sample

It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (0.114-30 mmol/L).

The recommended dilution factor for different samples is as follows (for reference only):

Sample type Dilution factor
Rabbit plasma 1
Rat serum 1
Rat plasma 1
Human serum 1
Human saliva 1
Human milk 1
Human urine 30-60
Mouse urine 30-60

 

Note: The diluent is normal saline (0.9% NaCl) or PBS (0.01 M, pH 7.4).

 

Detection Principle

Urea can be decomposed into ammonia ion and carbon dioxide by urease. Ammonia ion can react with phenol chromogenic agent and form a blue substance in alkaline medium, and the production of the blue substance is proportional to the urea content which can be calculated with the colorimetric assay at 580 nm.

 

Kit Components & Storage

Item Component Size 1
(50 Assays)
Size 2
(100 Assays)
Storage
Reagent 1 100 mmol/L Urea
Standard
1 mL × 1 vial 2 mL × 1 vial 2-8°C, 12 months
Reagent 2 Enzyme Stock
Solution
0.05 mL × 1 vial 0.1 mL × 1 vial 2-8°C, 12 months, shading light
Reagent 3 Enzyme Diluent 15 mL × 1 vial 30 mL × 1 vial 2-8°C, 12 months
Reagent 4 Chromogenic
Agent
60 mL × 1 vial 60 mL × 2 vials 2-8°C, 12 months, shading light
Reagent 5 Alkaline NaClO 60 mL × 1 vial 60 mL × 2 vials 2-8°C, 12 months, shading light

 

Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

Technical Data:

Parameter:

Intra-assay Precision

Three human serum samples were assayed in replicates of 20 to determine precision within an assay. (CV = Coefficient of Variation)

Parameters Sample 1 Sample 2 Sample 3
Mean (mmol/L) 3.50 16.80 22.40
%CV 5.2 4.6 4.0


Inter-assay Precision

Three human serum samples were assayed 20 times in duplicate by three operators to determine precision between assays.

Parameters Sample 1 Sample 2 Sample 3
Mean (mmol/L) 3.50 16.80 22.40
%CV 5.4 4.8 3.9


Recovery

Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 102%.

Sample 1 Sample 2 Sample 3
Expected Conc. (mmol/L) 8.5 18.5 26
Observed Conc. (mmol/L) 8.4 19.4 26.5
Recovery rate (%) 99 105 102


Sensitivity

The analytical sensitivity of the assay is 0.114 mmol/L. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.


Standard Curve

As the OD value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only.

Concentration (mmol/L) 0 4 8 12 16 20 30
Average OD 0.007 0.068 0.13 0.207 0.268 0.342 0.519
Absoluted OD 0 0.061 0.123 0.2 0.261 0.335 0.512